3.1 Turning on device
a. Plug the power cable into a power outlet
(Fig. 3.3) of the correct voltage and frequency
as defined on the manufacturer's technical label
b. Press the power switch
(I/
O
)
(item 1 in Fig.2.1)
to the ON position
(I)
. The green indicator light
flashes until concentration is achieved.
TURN THE ANESTHESIA MACHINES OXYGEN
FLOW METER ON TO A MINUMUM FLOW RATE
OF 0.25 LPM
A "NO FLOW" ALARM WILL SOUND
IF THE ANESTHESIA MACHINE FLOW METER
ISN'T OPEN
1. Filter / Silencer
2. Cabinet filter
3. Ventilation grill
Note: Shown with grate removed
4. CLEANING - MAINTENANCE
Only the outside of the concentrator
is to be
cleaned. Use a damp sponge or cloth with
water only.
Acetone, solvents or any other inflammable
products
must not be used
. Do not use abra
-
sive powders.
Cabinet air filter
(Item 1 Fig.4.1) must be
cleaned in warm water and household detergent
weekly or after approximately 100 hours of use.
Dry before reinstalling. More frequent cleaning
is recommended in dusty environments.
Fine filter
(Item 2 Fig.4.2) remove the right
side filter grate (Item 3 Fig.4.2) The fine intake
filter should be replaced annually or every
2000 hours of use. More frequently dusty /
dirty environments.
P/N OC8000-2
Fig. 4.1
1
2
3
Fig. 4.2
3.2 Turning off device
At the end of the usage, press the
I/
O
Switch
to place it in the
O
(OFF) position to stop the
device. The oxygen enriched air flow continues
for approximately one minute after the device
is stopped.
Note:
After turning the unit off, the user must
wait 3-5 minutes before turning it back on.
System pressure must dissipate before the unit
will properly restart.
FILTER MAINTENANCE