Appendix 1: PGA 3510 Common Modbus Registers
This appendix covers Modbus registers most commonly accessed by PGA 3510 users. These are Modbus
registers 200 through 227 and their descriptions. There are other Modbus registers used by the PGA
3510. If you need information on other registers, call SSi at (513) 772-0060.
Register Location
Parameter
Description
200
co_data
Measured CO value
201
co2_data
Measured CO
2
value
202
o2_data
Measured 0
2
value
203
ch4_data
Measured CH
4
value
204
Bench_off
Turn off bench, pump, and sample
205
IR_carbon
IR %C
206
IR_COF
Suggested COF (CO Factor)
207
IR_PF
Suggested PF (Process Factor)
208
IR_MV
IR equivalent millivolts
209
probe_MV
Probe millivolts
210
probe_COF
Probe COF
211
probe_temperature
Probe temperature
212
probe_PF
Probe PF
213
probe_carbon
Probe %C
214
pump_run_min
The total time in minutes that the
pump has run since the last time
memory was cleared or a manual
reset took place. This value, plus
pump_run_wks, will provide the total
amount of time the pump has been
running. When this number reaches
10080, pump_run_wks is increased by
1, and pump_run_min is reset to zero.
215
pump_run_wks
The total time in weeks that the pump
has run since the last time memory
was cleared or a manual reset took
place. This value, plus pump_run_min,
will provide the total amount of time
the pump has been running.
216
pump
Pump status
217
pump_run_now
The time in minutes that the pump is
running (reset to zero if pump is
turned off). Maximum value is 30000. If
pump_run_now is greater than
pump_run_max, then pump_run_max
is set to pump_run_now.
218
pump_run_max
The longest time in minutes of a pump
session. If pump_run_now is greater
than pump_run_max, then
pump_run_max is set to
pump_run_now.
219
calib_status
Calibration status
220
IR_temperature
IR temperature
221
output_percent
Control output in percent
222
sample_flow
Sample flow