
6.5 Multi-Cal
Enter setup of multi-points calibration to set the number of calibration points. This function
only applies to pH mode.
Enter ”Temperature” Setup
Press or to select the number of calibration points. There are
1~3 points to choose from. The factory default is 2 points. When the
calibration reaches the number of setting points, the calibration
procedure will automatically finish and display the calibration result.
ġ ġ ġ ġ
Press to confirm it.
Press to confirm it.
лр
Содержание PC-3310
Страница 1: ...PC 3310 Intelligent pH ORP Transmitter Operation Manual...
Страница 55: ......