3.4 Installation of accessorial transmitter PH-300T
ġġ
(Optional)
Accessorial pH/ORP transmitter, PH-300T, is mainly installed on the electrode protective pipe, but
also can apply wall mounting and pipe mounting. For long distance transmission (100m), if
PC-3110 is more than 30m far away from the electrode, PH-300T accessorial transmitter is
recommended to avoid the attenuation of electrode signal, and for the convenience of onsite
observation, measurement, and calibration.
Illustration of wall mounting type
1. Combine the sun shield and PH-300T transmitter by round stainless steel screws 4x15
2. Fix item 1 combination on wall by self-tapping screws
Illustration of pipe mounting type
1. Fix immersive holder(B) into U-shaped clamp(A)
2. Combine the item 1. combination with PH-300T(C) by Round stainless steel screws 4
ǘ
15
A------Sun shield of PH-300T
B------PH-300T transmitter
C------Round stainless steel screws 4x15
D------Self-tapping screws(by customers)
A------U-shaped clamp
B------Immersive holder
C------PH-300T transmitter
D------ Round stainless steel screws 4
ǘ
15