7.4.5 TECH, NIST buffer Calibration
The procedure below is two points calibration of TECH buffer. (The procedure
is same as NIST buffer mode.) First, enter the setup of Multi-points calibration and
set the number of calibration point for 2. (See chapter 6.4 Multi-Cal) Then, go to
Calibration menu and select TECH mode. Operate the instrument as follow
procedure diagram.
Press or to
change the page of
calibration result display
Under MTC(Manual
Temp. Compensation)
temperature mode,
press
or
to adjust temperature.
If under ATC(Auto
Temp. Compensation)
temperature mode,
(PTC or NTC), the
temperature value is
obtained by itself.
Press to decide to
simply make a single
point calibration.
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
Under MTC(Manual
Temp. Compensation)
temperature mode,
press
or
to adjust temperature.
If under ATC(Auto
Temp. Compensation)
temperature mode,
(PTC or NTC), the
temperature value is
obtained by itself.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Obtain the pH value of
first point buffer
solution.
Obtain the pH value of
second point buffer
solution.
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
44