6.8 Relay 1
Enter setup of Relay 1. Select the item to turn on or turn of the relay 1 function. If you
select to turn on the relay 1, then select for using relay 1 as “Hi set-point” alarm or
“Low set-point” alarm. Set the value of set-point (SP) and dead-band (DB). The range
for set-point is -2.00~16.00pH/-1999~1999mV; while the range for DB is 0.00~2.00pH
/0~200mV.
Note: When PID is set as “Pulse Length” or “Pulse Frequency” control, and the
control way is set as “Direct”, the transmitter uses Relay 1 as PID output control
device. At the time, the relay 1 setting function is turned off.
Press or to adjust
dead-band(DB) value
Press
or
to select to activate
REL1 or not. If
not, the guide
menu goes to setup
of Relay 2.
Press to confirm it.
Enter “Relay 2
” Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or
to select use REL1
as Hi point or Lo
point alarm.
Press
or
to select to activate
a test of REL1 or
not. If “ON”, the
relay 1 is in action,
and the “ACT”
indication light
will light up.
Press or to adjust
set-point(SP) value
28