Due to the manufacturing process, minor molding blemishes may be present on the protective armour. These
blemishes will not impact the functionality of the armour or the monitor and are not considered defects of the
product.
Cleaning the Masimo SpO
2
Accessories
WARNING:
Do not immerse the connector end of the sensor cable as this may damage the sensor.
NOTE:
Cleaning instructions for the M-LNC Patient Cable, YI AH Sensor, and TFI AH Sensor may be found in the
directions for use (DFU) of those items.
Cleaning the AccuVet Temperature Probe
Clean the probe before and after each use making sure to remove all bioburden. Brief immersion of the probe in
cleaning solution is acceptable except for the probe connector which must not be immersed. Disinfection of the
probe and its connecting cable can be done with a soft cloth saturated with 70% isopropyl alcohol or 10% (1:10)
solution of chlorine bleach and tap water. Afterwards, wipe the probe with a soft cloth dampened with clean water.
Allow the probe and cable to dry thoroughly before use.
Cleaning the AccuVet SpO
2
Y Sensor
WARNING:
DO NOT immerse the SpO
2
probe in any fluid, or attempt to clean the unit with any liquid detergents or
solvents. Do not sterilize. This may cause an electrical hazard.
Clean the sensor and clip before and after each use. Clean the clip, sensor, and connecting cable with warm soapy
water or 70% isopropyl alcohol. Allow the sensor, clip, and cable to dry thoroughly before use.
Cleaning the AccuVet SpO
2
Reflectance Sensor
CAUTION:
Do not sterilize by
irradiation, steam, or ethylene oxide.
Clean the sensor before and after each use making sure to remove all bioburden. The sensor may be surface-cleaned
by wiping it with a soft cloth saturated with a solution such as 70% isopropyl alcohol. If low level disinfection is
required, use or 10% (1:10) solution of chlorine bleach and tap water. Do not use undiluted bleach (5%-25% sodium
hypochlorite) because permanent damage to the sensor could occur. To clean or disinfect the sensor:
1.
Saturate a clean, dry gauze pad or soft cloth with the cleaning solution. Wipe all surfaces of the sensor with
the gauze pad.
2.
Saturate another clean, dry gauze pad or soft cloth with sterile or distilled water. Wipe all surfaces of the
sensor with the gauze pad.
3.
Allow the sensor and cable to dry thoroughly before use.
Cleaning the AccuVet SpO
2
Extension Cable
Clean the connecting cable with warm soapy water or 70% isopropyl alcohol. Allow the cable to dry thoroughly before
use.
10.3 Preventative Maintenance
WARNING:
Do not disassemble the unit. There are no user serviceable parts except for the battery. Refer to qualified
service personnel.
System Self Checks
The Vet30E performs a range of system checks during normal operation. If the monitor detects a problem, it will
display a status message recommending a troubleshooting action or to contact SunTech Customer Service.