Setting up a Remote Site
Turn on the modem by depressing the On/Off button on the rear of the modem. Check that the
power light on the front of the modem illuminates to ensure a proper connection. With a phone
connected to the modem, (optional), you can pick up the phone and check that you hear a dial
tone to ensure correct connections of the phone line.
Note
: the Oscar Express system is designed for analog phone lines
ONLY
. Data transfers through
office PBX or other digital phone systems are not expected to be compatible with the Express
modem data transfer.
Voice communication options
When preparing to do a modem data transfer with the remote site, the central terminal operator
needs to communicate with the remote site to:
•
Identify which remote site is about to send data
•
Identify the remote user they are speaking to
•
Identify the patient for the data about to be sent
•
Determine if the Oscar 2 monitor is to be programmed for a new patient study and what
parameters are needed
•
Ensure that the data transfer was successful after completion
You can use one of two remote site setups illustrated on the following pages to do this. Option 1 is
highly recommended
, as it minimizes the chance of ambient background noise interfering with
the modem communication. However, if the remote site only has one phone line, Option 2 will
work. Regardless of the method used, data communication begins with a person-to-person call to
the central terminal.