The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure
that it is used in such an environment.
16
Cables
Length
(
m
)
Whether to block
Note
USB Cable
1
No
-
Guidance and Manufacturer’s declaration – electromagnetic emissions
This device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should assure that
it is used in such environment.
Electromagnetic emission IEC 60601-1-2
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
This device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
This device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/flicker-
emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s declaration – electromagnetic immunity
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment
– guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6 kV contact
±8kV air
±6 kV contact
±8kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%.