Nevertheless, special precautions need to be observed: 

1.NOTE The EMISSIONS characteristics of this equipment make it suitable for 

use in a residential environment (for which CISPR 11 class B is normally 

required).

2.WARNING:The use of accessories and cables other than those specified by 

Sunset, with the exception of cables sold by Sunset as replacement parts for 

internal components, may result in increased emission or decreased immunity 

of the device.

3.WARNING:The medical devices should not be used adjacent to or stacked 

with other equipment. In case adjacent or stacked use is necessary, the 

medical device should be observed to verify normal operation in the 

configuration in which it will be used.

4.WARNING: PORTABLE RF communications equipment (including peripherals 

such as antenna cables and external antennas) should be used no closer than 

30cm (12 inches) to any part of [ME QUIPMENT or ME SYSTEM], including cables 

specified by the MANUFACTURER. Otherwise,degradation of the performance 

of this equipment could result.

5.WARNING:Do not use mobile (cellular) telephones and other devices(such as 

MRI,diathermy,electrocautery,RFID and electromagnetic security systems) 

which generate strong electrical or electromagnetic fields, near the medical 

device. This may result in incorrect operation of the unit and create a 

potentially unsafe situation. Recommendation is to  keep a minimum distance 

of 7m. Verify correct operation of the device in case the distance is shorter.

15

to electromagnetic interference from other devices. Electromagnetic 

interference may result in incorrect operation of the medical device and create 

a potentially unsafe situation.

Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) 

with the aim to prevent unsafe product situations, the IEC60601-1-2 standard 

has been implemented. This standard defines the levels of immunity to 

electromagnetic interferences as well as maximum levels of electromagnetic 

emissions for medical devices.

Medical devices manufactured by Sunset Healthcare Solutions (Sunset) 

conform to this IEC60601-1-2 standard for both immunity and emissions.

Содержание NEB402

Страница 1: ...Instruction Manual NEB402 Handheld Mesh Nebulizer IFU 00059 REV 1 01 2022 10 12 Manufactured for Sunset Healthcare Solutions 180 N Michigan Ave Ste 2000 Chicago IL 60601...

Страница 2: ......

Страница 3: ...very much for choosing our Handheld Mesh Nebulizer Please read this user manual carefully before using so that you can use it safely and correctly Please keep this user manual properly in place for fu...

Страница 4: ...5 5 Product Technical Parameters 6 6 User Instructions 7 7 Cleaning Disinfection 10 8 Storage Conditions 12 9 Trouble Shooting 12 10 Note Warning 13 11 EMC Declarations 14 12 Signs Symbols 21 13 After...

Страница 5: ...device if it is damaged or becomes wet The device is a Vibrating Mesh Nebulizer System designed to aerosolize liquid medication for inhalation therapy by the patient The device may be used with pedia...

Страница 6: ...ion Rate 0 25ml min 0 9ml min Particle Size MMAD 5 m Medication Cup Capacity 10ml Max Vibration Frequency 110kHz 10kHz Weight 100g with batteries Size Length 38 5mm Width 53 4mm High 97 7mm Working En...

Страница 7: ...rable period may vary depending on usage environment Main Unit 2 years Top Cover 2 years Medication Cup 2 years Medication Cover 2 years Mask PVC 1 year The nebulizer has a USB cable for charging it d...

Страница 8: ...the device every 2 days to ensure normal use Continue charging the device at least once per month when not in use for more than one month The rechargeable battery should not be replaced by the user o...

Страница 9: ...elow and click ON OFF switch to power on Please hold the nebulizer to keep liquid medication in contact with the vibrating mesh and the stream of inhalation Please see Figure 6 below Figure 5 Connect...

Страница 10: ...medication cup and mask or mouthpiece We recommend one device be used and reused for a single person The device is not provided sterile 10 Step Four Figure 7 Pouring out the Residual Liquid Medicatio...

Страница 11: ...owing the cleaning instructions after each use will prevent any remaining medication in the bottle from drying adhering to the mesh cap and resulting in the device not nebulizing effectively Please do...

Страница 12: ...ditions 9 Trouble Shooting Item Trouble Possible cause solution 1 Does not work when turned on Check if there is enough power Check if there is enough medication Check if the medication cup has been f...

Страница 13: ...non conforming adapter may cause equipment damage Please refer to the user manual when problems appear and contact the service center for maintenance Please clean and disinfect the medication cup when...

Страница 14: ...may cause Bronchospasm The device can t be used on an aircraft One device is solely for single patient use and can be reused by a single patient One device is used for single medication Do not use mob...

Страница 15: ...ce of this equipment could result 5 WARNING Do not use mobile cellular telephones and other devices such as MRI diathermy electrocautery RFID and electromagnetic security systems which generate strong...

Страница 16: ...function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B This device is suitable for use in all estab...

Страница 17: ...lity should be that of a typical commercial and or hospital environment If the user of this device requires continued operation during power mains interruptions it is recommended that the device be po...

Страница 18: ...3V rms 3V m Portable and mobile RF communica tions equipment should be used no closer to any part of this device including cables than the recommend ed separation distance calculated from the equation...

Страница 19: ...should be less than 3 V m 19 NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorpt...

Страница 20: ...bove the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter...

Страница 21: ...mful substances and element Protected against solid foreign objects of 12 5 mm and greater Protection against vertically falling water drops when ENCLOSURE tilted up to 15 Date of manufacture Keep awa...

Страница 22: ...push Operation indicator Storage and delivery temperature limits Storage and delivery humidity limitations Storage and delivery atmospheric pressure limitations Micro USB plug interface Press inside t...

Страница 23: ...ed accordingly 2 Please contact us to obtain warranty service Warranty conditions To obtain the warranty service please present this warranty card and fill out the related content The company provides...

Страница 24: ...riginal parts or accessories The manufacturer does not take responsibility for the buyer or third parties for any damage or loss intentionally or unintentionally caused by improper use On the request...

Страница 25: ...if returning machine to maintenance The warranty period for the unit is 1 year The warranty for the medication cup is 6 months During the warranty period Sunset will decide to repair or replace damage...

Страница 26: ...mful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful...

Страница 27: ......

Страница 28: ......