Sunrise Medical Switch-It CTRL+5 Скачать руководство пользователя страница 7

7

247780   Switch-It CTRL+5 with Assignable Buttons   Rev.A

INSTALLATION

Install the CTRL+5 in a location that is easy for the intended user to 

access.

Only authorized Sunrise Medical dealer should install this product.

INDICATORS

The behavior of the indicators is dependent on the seating module in the 

system.

1. 

With an Intelligent Seating Module (ISM) the indicators, Figure 6,will all 

be green or yellow.

• 

Yellow indicates a drive inhibit is active and green indicates there is 

no drive inhibit.

• 

All white indicators meant the module has not authenticated with the 

powerchair.

2. 

With a Complex Seating Module (CxSM) the indicators, Figure 6, are 

linked to the button below them.

• 

When a button is assigned to any other option but seating, the light 

will be solid white.

• 

When a button is assigned to a seating Memory axis, and that axis is 

running, the light will be blinking green.

• 

When a button is assigned to a seating Memory axis and a drive 

inhibit occurs the LED will change to the drive inhibit color once the 

axis is stopped.

• 

When a button is assigned to a seating Memory axis other than a 

Memory axis the drive inhibit color change will happen when the 

inhibit happens.

• 

The LED color will be Red for a lockout drive inhibit.

• 

When a button is assigned to any seating axis and no actuator 

channel within that axis has called for a drive inhibit the LED color will 

be Green.

Fig. 6

Fig. 7

Fig. 8

PROGRAMMING ASSIGNABLE BUTTONS

1. 

Changing the assignable buttons of the Ctrl+5 should only be done 

by an authorized Sunrise Medical dealer.

2. 

Connect powerchair to PC Programmer using a Programming 

Dongle.

3. 

Read Chair into the programmer by clicking on the blue down arrow 

in the toolbar, shown at right. 

4. 

Open the Switch-It folder and the module you wish to change. 

Figure 7 shows Button 1 of module 1 is assigned to Speed Up and 

Speed Down.

5.  To change this double click on the current assignment. In this case 

you would double click on the “Speed Up” text. This brings up the 

window as shown in Figure 8.

6. 

Now select the function you need that button to do by clicking the 

checkbox, then click OK.

7. 

Continue this process until all buttons have the desired function

8. 

Now click the upload button, shown at right.

9. 

Now power cycle the chair. Your new assignments are now updated.

Indicator Window

Содержание Switch-It CTRL+5

Страница 1: ...Switch It CTRL 5 with Assignable Buttons 000247780 01 Owner s Manual...

Страница 2: ...ct a qualified clinician or supplier otherwise injury or damage may occur WARNING This device can be affected by Electro Magnetic Interference EMI and Radio Frequency Interference RFI Radio wave sourc...

Страница 3: ...t initiative Sunrise Medical reserves the right to change specifications and design without notice This user manual This user manual will help you to use and maintain your product safely Dealer This o...

Страница 4: ...nings Safety Temperature WARNING If you do not use the Ctrl 5 ensure that it is not exposed to direct sunlight for lengthy periods of time Certain parts of the Ctrl 5 such as the keypad the cable and...

Страница 5: ...e sure that you dry all parts of the Ctrl 5 prior to use Be sure to clean your Ctrl 5 regularly This will help you find loose or worn parts If you find loose damaged or worn parts replace before use C...

Страница 6: ...ly 4 25 108 mm W x 2 5 63 5 mm H x 1 25 31 75 mm D USB port on the Ctrl 5 provides up to 2 0 Amps C Compatibility Penny Giles R Net Electronics required Power wheelchair must have assignable buttons f...

Страница 7: ...ssigned to a seating Memory axis other than a Memory axis the drive inhibit color change will happen when the inhibit happens The LED color will be Red for a lockout drive inhibit When a button is ass...

Страница 8: ...hair off 2 Plug Ctrl 5 back in and turn chair on Issue New button assignments are not loading on the device Solution 1 Unplug Ctrl 5 and turn chair off 2 Power cycle turn off and turn back on the chai...

Страница 9: ...is not limited to the following parts where fitted batteries armrest pads upholstery tyres brakes shoes ferrules etc b Any overloading of the product please check the EC label for maximum user weight...

Страница 10: ...10 247780 Switch It CTRL 5 with Assignable Buttons Rev A...

Страница 11: ...11 247780 Switch It CTRL 5 with Assignable Buttons Rev A...

Страница 12: ...rillonnerie 17 Rue Micka l Faraday 37170 Chambray L s Tours Tel 33 0 247554400 Fax 30 0 247554403 www sunrisemedical fr Sunrise Medical S r l Via Riva 20 Montale 29122 Piacenza Italia Tel 39 0523 5731...

Отзывы: