T R O U B L E S H O O T I N G
TROUBLESHOOTING CHART A
Visible Alarm
Audible Alarm
Compressor
Power Light
OFF
OFF
ON
ON
Other Symptoms
Possible Cause
Possible Remedy
Pulsating air noise
Intake filter not in place or defective
Check filter and replace if necessary
Compressor intake hose disconnected
Reconnect hose
Excessive noise
Loose or defective motor mounts
Replace motor mounts
Defective compressor
Replace compressor
Defective cooling fan
Replace cooling fan
Fluctuating oxygen flow
Occluded humidifier
Clean or replace humidifier
Use of improper humidifier
Use only a bubble-type humidifier
Occluded filters
Clean or replace filters
Occluded or defective cannula and tubing
Detach cannula from oxygen delivery tubing. If
proper flow is not attained, check tubing for
kinks or other obstructions. Clean or straighten
as required or replace tubing if necessary.
Use of excess oxygen tubing
The unit is designed to deliver 5 lpm with a
cannula on 50 feet (15 meters) of approximately
5/32” (4 mm) inside diameter tubing. Smaller
diameter tubing or the addition of any other
flow restriction may prevent obtaining the
desired flow rate.
Defective flow meter
Replace flow meter
Leak in system
Check for leaks in all hoses and fittings
Defective compressor
Replace compressor
Defective compressor reed valve
Replace compressor reed valve
Defective check valve
Replace check valve
Pressure regulator not adjusted
Adjust or replace pressure regulator
properly or defective
Little or no oxygen flow
Flow meter not adjusted properly
Adjust flow meter
Hose disconnected to flow meter
Reconnect hose
Oxygen delivery tubing is kinked or blocked
Straighten tubing or remove obstruction
Occluded humidifier
Clean or replace humidifier
Low oxygen concentration
Leak in system
Check for leaks in all hoses and fittings
Defective sieve bed check valve
Replace check valve
Defective compressor reed valve
Replace compressor reed valve
Defective compressor
Replace compressor
Rotary valve not operating correctly
Replace valve
Occluded filters
Clean or replace filters
Contaminated sieve beds
Replace sieve beds
Audible alarm does not
Unit has not been used for an extended
Allow unit to run for 20 minutes and retry
sound during power failure
period of time.
NOTE
: If the concentrator
has been unused for an extended period,
the unit must run several minutes before
the power fail alarm will activate.
Defective PC board
Replace PC board
Defective power switch
Replace power switch
Defective wire harness
Replace wire harness
Audible alarm does not sound when Defective PC board
Replace PC board
unit is turned “On”
Pressure relief valve activated -
Defective PC board
Replace PC board
“popping” sound
Defective rotary valve
Replace valve
Service Required light does not
Defective PC board
Replace PC board
illuminate when unit is turned “On”
Defective light
Replace light
PC board connectors not properly latched
Be sure tabs are pushed completely into place
Содержание DeVilbiss 515ADS
Страница 43: ...43 LT 1928 F I G U R E S D I A G R A M S A N D V I E W S Figure 14 Pneumatic Diagram...
Страница 44: ...44 LT 1928 F I G U R E S D I A G R A M S A N D V I E W S Figure 15 Wiring Diagram 115 VOLT UNITS 230 VOLT UNITS...
Страница 92: ...92 LT 1928 F I G U R A S D I A G R A M A S Y V I S T A S Figure 14 Diagrama Neum tico...
Страница 93: ...93 LT 1928 F I G U R A S D I A G R A M A S Y V I S T A S Figura 15 Diagrama de Cableado...
Страница 99: ......