800-433-2797 www.Sun-Med.com
User Guide
To Use the Nebulizer with a Mask
1. Remove the mouthpiece and place in respiratory equipment bag,
in case you need it later.
2. Remove the inspiratory valve and place on the side elbow port. (A) .
To ensure a proper fit, push and turn the valve into the side port .
3. Keep the nebulizer upright. Add the prescribed medication through
top of the reservoir tower (B).
4. Insert mask connector into the top of the reservoir tower (C).
• If not using a Salter Labs aerosol mask, use the universal mask
adapter included in package to connect the mask to the nebulizer.
5. Set the flow rate from 5 LPM to 8 LPM (25 PSI to 52 PSI)
to begin the treatment.
6. Apply the mask to the patient
Alternative Mask Placements
If the patient is unable to sit upright or is in a semi-fowlers position,
attach the mask to the nebulizer’s side port.
Setup A
1. Connect mask to the side port.
Setup B
1. Additional equipment is required. Gather a standard 15 mm ID / 22 mm
OD adapter and a 6 inch corrugated aerosol tubing with 22 mm ID.
2. Attach adapter to side port.
3. Attach one end of corrugate tubing to the adapter
and the other end to the aerosol mask.
NebuTech® 8960 Disposable Nebulizer Cont.
Disposable | Single Patient Use
B
A
C
Setup A
Setup B
1. ALL GRAPHICS UPDATE