26
4.
Data calibration mode: switch between manual and auto.
5.
Lens type: select lens type from drop-down list.
6.
A-scan tested data display : show at most 10 groups of measurement data and mean
number, standard deviation.
7.
Clear: Click to clear current patient A scan test data.
8.
Ignore: Select a set of A scan test data,click “Ignore” then the set of data don’t count
in average value , standard deviation and IOL calculation. Click the “ignore ”again,the set
of data recover to normal , and participate in calculation. If you need ignore multi-group
data, select one set , click ignore , then select another set , click ignore.
9.
Delete: Delete selected test data.
10.
Enter /shift : under manual examination mode, you can switch among A/L/V
calibration point by pressing it.
11.
Left & Right move icon: move the calibration point under manual mode
12.
Start/Stop : start collecting data or stop collecting data by pressing it.
13.
A-scan echogram display area.
14.
Tissue velocity: show the tissue velocity under current mode.