Consult the MSDS that accompanies each standards set to identify proper 

disposal techniques.

If  you  experience  an  issue  with  the  printed  circuit  assembly  (PCA)  that 

contains a lithium battery, please contact SUEZ for return instructions. 

Whenever changing the lithium battery, observe the correct polarity. Dispose 

of used batteries according to the manufacturer’s instructions and local 

disposal requirements.
To protect the Analyzer from damage, always drain the water from the 

Analyzer prior to shipping.

Warnings    

 (continued from page 1)

This symbol

 

on the instrument indicates that the user should refer to 

the manual for operating instructions. 

The vial sparging assembly contains sharp needles designed to pierce 

the septa of sample vials. Do not put fingers or inappropriate materials 

into the vial sparging assembly.

For continued protection against fire hazard, replace fuse with same 

type and rating.

This symbol indicates the protective earth terminal (ground) for the 

Analyzer.

This symbol indicates that to comply with European Union Directive 

2002/96/EC for waste electrical and electronic equipment (WEEE), 

the Analyzer should be disposed of separately from standard waste.

This symbol on the instrument indicates that the user should refer to 

the manual for instructions on filling the DI water cartridge, as found 

in the Installation and Maintenance chapters.

To protect against accidental exposure to ultra-violet radiation, do not operate 

the UV lamp outside of its protective housing.

This is a Safety Class I product. It must be must be attached to a grounded 

power source.

If this instrument is used in a manner not specified by SUEZ, the protection 

provided by the instrument may be impaired.

Always stop analysis before turning off or unplugging the Analyzer.

Make sure the DI water cartridge is full, particularly when running samples 

with high TOC or high salt concentrations. Always “clean-up” the Analyzer by 

running low-TOC DI water after running high TOC or salt samples.

This is a Class A product. In a domestic environment, this product may cause 

electromagnetic interference in which case the user may be required to take 

adequate measures to correct the interference.

To avoid false TOC readings and possible damage to the Analyzer, always 

make sure the sample inlet is open and the DI water cartridge is filled before 

starting analysis.

The glass window in the UV reactor indicates proper lamp operation and 

should not be viewed too closely (closer than 6 inches) or for extended 

periods (greater than 1 minute). Protective eyewear is recommended.

This symbol on the Sievers 860 TOC  Analyzer indicates that the 

product does contain restricted substances included in China RoHS II. 
Please refer to the following table.

5

6060 Spine Road

Boulder, CO 80301-3687 USA

total organic carbon

1

 

Linear Range 
Precision 
Accuracy 
Analysis Time 
Sample Flow Rate (nominal)
 
Sample pH

0.3 ppb – 1.5 ppm TOC 
0.1 ppb or 1%, whichever is greater 
± 5% at 500 ppb sucrose standard 
6 minutes 
Analysis mode:  0.25 mL/min 
Fast Flush (between samples): 0.5 mL/min 
<3

analyzer specifications 

Power Requirements 
Fuses 
Ambient Temperature 
Normal Operating Environment 
Maximum Relative Humidity 
Maximum Altitude 
Sparge Gas Inlet Pressure
Sparge Gas Flow Rate 
Calibration Stability 
Outputs 
Installation/Overvoltage Category 
Safety Certifications
 
Pollution Degree 
Display 
Size 
Weight 

100–240 ±10% VAC, 60 watts, 50/60 Hz 
T1.6 A, 250 VAC (SloBlo), size 5 x 20 mm (appliance inlet) 
10 °C to 40 °C 
Intended for indoor use only 
Up to 95%, non-condensing 
3,000 m (9,843 ft) 
30 psig to 50 psig 
120 mL/min to 150 mL/min 
Typically stable for 12 months 
USB, Ethernet port 
II 
CE, ETL listed. Conforms to UL Std. 61010-1. 
Certified to  CSA C22.2 No. 61010-1 

Color LCD 
H: 43.0 cm (16.9 in); W: 21.5 cm (8.5 in); D: 42.0 cm (16.5 in) 
10.5 kg (23.0 lb) 

Sievers Autosampler RT12 specifications

Capacity 
Power Supply 
Ambient Temperature 
Normal Operating Environment 
Maximum Relative Humidity 
Maximum Altitude 
Size 
Net Weight 
Safety Certification 

Twelve (12) 40-mL vials 
Power supplied by the Sievers 860 Laboratory TOC Analyzer 
10 °C to 40 °C 
Intended for indoor use only 
Up to 95%, non-condensing 
3,000 m (9,843 ft) 
H: 36.8 cm (14.5 in); W: 23.1 cm (9.1 in); D: 25.4 cm (10.0  in) 
3.9 kg (8.6 lb), including the empty sample rack 
CE 

1  

Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.

DQS 79000-03 EN Rev. A

Содержание Sievers 860

Страница 1: ...rotection when using compressed gases to avoid possible eye injury If the Analyzer is used with compressed gases the pressure of the gas at the gas inlet must not exceed 200 psig 1379 kPa Use only CO2 free air or nitrogen Use only regulators designed for use with the selected gas Store compressed gases in accordance with all safety codes Before connecting the Sievers Autosampler RT12 to the Analyz...

Страница 2: ...idge 7 Slide the water bottle nozzle into the inlet hole Fill the cartridge until water reaches the FULL line 8 Replace the rubber inlet cover Prime the DI Water Pump 1 Connect the power cord to the Analyzer 2 Turn the Analyzer On The DI water pump will start in 20 30 seconds 3 Check for large air bubbles in all the tubing that connects to the DI water cartridge If large air bubbles are present ta...

Страница 3: ... and the base of the rack is flush with the Sievers Autosampler RT12 Figure 6 Step 10 D Install the sample tubing 1 Locate the end of the sample tubing with the PEEK fitting with yellow ferrule Attach it to the Sample port on the Autosampler and then tighten finger tight Figure 7 2 Route the sample tubing from the Autosampler to the Analyzer Secure the tubing against the Analyzer to ensure that th...

Страница 4: ...oes not make contact with the needle guide Reduce the Needle Down Depth if either of these conditions are not met 5 If adjustment is needed in the Needle Down Depth area use the Depth up and down arrows to adjust the needle depth 6 Click the Set button to store the new setting 7 Check your results by disconnecting DataPlus from the Analyzer To do this press the Disconnect button 8 Cycle power on t...

Страница 5: ...ich case the user may be required to take adequate measures to correct the interference To avoid false TOC readings and possible damage to the Analyzer always make sure the sample inlet is open and the DI water cartridge is filled before starting analysis The glass window in the UV reactor indicates proper lamp operation and should not be viewed too closely closer than 6 inches or for extended per...