Sievers 860 Laboratory

TOC Analyzer

and Sievers Autosampler RT12

Quick Start Guide

IMPORTANT NOTICE

 

For more detailed instructions, download the 

Sievers 860 

Laboratory TOC Analyzer Operation and Maintenance 

Manual 

from our website.

Installation Instructions

 Step 1 

     

Unpack and inspect the Analyzer.

 Step 2 

     

Complete the identification records.

 Step 3 

Select a location for the Analyzer. Refer to the System   

Specifications chart.

Place the Analyzer on a clean, unobstructed surface that can 

accommodate the Analyzer’s dimensions and weight; for details refer to 

the "System Specifications." For proper heat dissipation, ensure that 

16  cm (6.3 in) is available at the rear and on both sides of the Analyzer. 

The Analyzer is designed to be operated right-side up, resting on the 

rubber feet.

Avoid direct sunlight and extreme temperatures; operating at elevated 

temperatures (greater than 40 ºC) prevents proper operation, and 

operating at low temperatures (10 ºC) can cause errors in the 

measurements.

 Step 4 

     Connect the compressed gas.

Attach the Analyzer to a source of dry, oil-free gas that contains less than 

1 ppb CO

2

, and which is pressurized from 30 to 50 psig (207 to 345 kPa). 

The preferred gas is purified air. Nitrogen is also a suitable gas if the IC is 

less than 1 ppm.

Refer to the "Warnings" section for important information about the 

use of compressed gas.

1.  Locate the 1/8" OD tubing with 1/8" SwageLok fitting for gas inlet 

line, 3 meters long, in the accessories kit.

Warnings

The vial sparging assembly contains a sharp needle

 designed to 

pierce the septa of sample vials. Do not put fingers or inappropriate 

objects into the vial sparging assembly.

To avoid false TOC readings and possible damage

 to the Analyzer, 

always make sure the DI water cartridge is filled before starting analysis.

The Sievers 860 Laboratory TOC Analyzer 

is not designed to sample 

in on-line mode from a pressurized water source.

The waste stream from the Analyzer is acidic 

and must be disposed of 

according to local regulations.

Secure all gas cylinders

 to an immovable structure or permanent wall.

Wear eye protection when using compressed gases

 to avoid possible 

eye injury.

If the Analyzer is used with compressed gases,

 the pressure of the 

gas at the gas inlet must not exceed 200 psig (1379 kPa). Use only 

CO

2

-free air or nitrogen. Use only regulators designed for use with the 

selected gas. Store compressed gases in accordance with all safety 

codes.

Before connecting the Sievers Autosampler RT12 to the Analyzer, 

or disconnecting the Austosampler from the Analyzer, always stop TOC 

measurements/protocols,

 

and turn Off power to the Analyzer. 

Do NOT manually rotate the sample rack

 while the Sievers 

Autosampler RT12 is connected to the Analyzer, even when the Analyzer 

is NOT powered.

Do not use the Sievers Autosampler 

RT12

 to sample flammable 

liquids. 

The Sievers Autosampler RT12 should not be in contact with 

flammable liquids, nor be used near flammable liquids.

The Sievers Autosampler 

RT12

 sample needle is sharp.

 Avoid 

contacting the needle assembly during analysis, and use caution when 

handling the needle assembly when the needle is at rest.

Avoid the possibility of the Autosampler needle hitting the 

sample rack

 and damaging the needle during the alignment 

process. Make sure to align the Autosampler needle by small 

increments (we suggest by one), and test the new alignment position.

Any  operation  requiring  access  to  the  inside  of  the Analyzer,  including 

installation of maintenance items, could result in injury. To avoid potentially 

dangerous shock, turn off power and, if possible, disconnect from the 

power supply before opening the Analyzer.

The UV lamp contains mercury and may be considered hazardous 

material in your local area. Dispose of these items in accordance with 

federal, state, or local government regulations.

Should the UV lamp become broken or damaged it should be handled 

in accordance with your organization’s toxic waste handling procedure 

and disposed of in accordance with federal, state, or local government 

regulations.

See page 5 of this document for additional important warnings information and 

product specifications.

ENGLISH

Water Technologies & Solutions

www.sieversinstruments.com

 

1

DQS 79000-03 EN Rev. A

Figure 1 - Inlet and Outlet Ports on Back of Analyzer

Gas Liquid Waste Outlet Port

Liquid Waste Outlet Port

Power Cord Receptacle

Power Switch

Gas Inlet Port

Ethernet Port

USB Port

Содержание Sievers 860

Страница 1: ...rotection when using compressed gases to avoid possible eye injury If the Analyzer is used with compressed gases the pressure of the gas at the gas inlet must not exceed 200 psig 1379 kPa Use only CO2 free air or nitrogen Use only regulators designed for use with the selected gas Store compressed gases in accordance with all safety codes Before connecting the Sievers Autosampler RT12 to the Analyz...

Страница 2: ...idge 7 Slide the water bottle nozzle into the inlet hole Fill the cartridge until water reaches the FULL line 8 Replace the rubber inlet cover Prime the DI Water Pump 1 Connect the power cord to the Analyzer 2 Turn the Analyzer On The DI water pump will start in 20 30 seconds 3 Check for large air bubbles in all the tubing that connects to the DI water cartridge If large air bubbles are present ta...

Страница 3: ... and the base of the rack is flush with the Sievers Autosampler RT12 Figure 6 Step 10 D Install the sample tubing 1 Locate the end of the sample tubing with the PEEK fitting with yellow ferrule Attach it to the Sample port on the Autosampler and then tighten finger tight Figure 7 2 Route the sample tubing from the Autosampler to the Analyzer Secure the tubing against the Analyzer to ensure that th...

Страница 4: ...oes not make contact with the needle guide Reduce the Needle Down Depth if either of these conditions are not met 5 If adjustment is needed in the Needle Down Depth area use the Depth up and down arrows to adjust the needle depth 6 Click the Set button to store the new setting 7 Check your results by disconnecting DataPlus from the Analyzer To do this press the Disconnect button 8 Cycle power on t...

Страница 5: ...ich case the user may be required to take adequate measures to correct the interference To avoid false TOC readings and possible damage to the Analyzer always make sure the sample inlet is open and the DI water cartridge is filled before starting analysis The glass window in the UV reactor indicates proper lamp operation and should not be viewed too closely closer than 6 inches or for extended per...

Отзывы: