8
3.2.
Movement module
The movement module is a unit that contains mechanism of manipulator, arm with XYZ
direction, probes and other necessary parts.
Communication system
cuvette
Temperature and
status
Sample
loading
Test
Clean
release
detach
move
Temp control
heat
cool
Status detection
Sample rack
Open alarm
XYZ move
loading probe
Loading pump
Magnetic bead
Optics
pipeline
Inflow/outflow
pump
3.3.
Sensor module
The sensor module can be divided into three parts including test channel, temperature
board and main board
3.3.1. Test channel use mechanical method which could drive the bead inside cuvette during
test cycle;
3.3.2. Temperature board which could provide 37
℃
to incubation area, test area and also
reagent area to incubate samples and reagents;
3.3.3. Main board could detect and measure the bead inside cuvette, and then calculate the
right result by the bead movement.
3.4.
Liquid flow module
The sample flow rate is monitored by measuring the micropipette which calibrated flow
restrictor
.
Liquid flow module consists of inflow pump, inflow valve, injectors, electronic valves,
probes, pipelines, outflow pump and some other parts. It ensures correct sample
amount of the whole system.
Содержание SF-8100
Страница 1: ...1 Beijing SUCCEEDER Technology Development Co Ltd SF 8100 Automated Coagulation Analyzer ...
Страница 9: ...9 ...
Страница 21: ...21 6 6 Unpack machine Step1 Cut off Straps Step2 Take the box cover off ...
Страница 22: ...22 Step3 Take out attachments as circled in photo ...
Страница 45: ...45 11 Circuit description map The layout of the SF8100 is illustrated in the circuit diagram as following ...