43
Connect motion control system error
1. Start VSPM, setup correctly.
2. Check whether Comport is being used.
4. Check RJ45 network cable,
5. RJ45 communication error
(
ping 192.168.1.8 to
check
)
6. Check connection of communication board
7. Replace main control board
Communication timeout
or communication error
Use “ comport expert” to troubleshoot
Cuvette running out or cuvette sensor failure
1.
No cuvette
2.
Mechanical error, microsensor cannot spring back
3.
Optocoupler broken
Liquid level sensor failure
1.
Enough reagent or sample
2.
Setup correct MAX depth
3.
Setup correct sensitivity
4.
Replace new liquid sensor cable
5.
Replace motor board
6.
Replace main control board
Reset error
1.
Replace optocoupler
2.
Locate optocoupler position
3.
Clean bearing and lifter-shaft or change it.
Consumable items has remaining balance
Not enough IC count, need to charge IC card.
Consumable items become 0
1.
Check communication connection (bad contact
USB cable broken, power shortage of USB port )
2.
Use 2to1 USB cable
3.
Re-open test software
Lack of reagent or special cleaning liquid
1. Setup reagent in software
2. Replace reagent
No calibration curve
1. Re-calibrate, coefficient>0.98
2. Check “ one valid” in calibration interface
Содержание SF-8100
Страница 1: ...1 Beijing SUCCEEDER Technology Development Co Ltd SF 8100 Automated Coagulation Analyzer ...
Страница 9: ...9 ...
Страница 21: ...21 6 6 Unpack machine Step1 Cut off Straps Step2 Take the box cover off ...
Страница 22: ...22 Step3 Take out attachments as circled in photo ...
Страница 45: ...45 11 Circuit description map The layout of the SF8100 is illustrated in the circuit diagram as following ...