EXACTA 1350-TD SERIES Operating Instructions
Sturtevant Richmont
P/N 857424 Rev
–
07/15/2020
7
CALIBRATION VERIFICATION
Calibration Verification Notes
1.
The following procedure assumes the tool is still wirelessly connected to the radio dongle. Refer to the SR Connect
Setup section to reconnect if needed.
2.
The accuracy tables for each model are located at the end of this document.
3.
Perform the procedure in the clockwise direction first, then in the counter clockwise direction.
Calibration Verification Procedure
1.
If the tester has not been zeroed in the CW direction, zero the tester and load the tool in the clockwise (CW) direction
to 100% capacity and then unload to 0% 3 times, remove the tool, and zero the tester.
2.
Click “Terminal”.
3.
If the tool is in Torque Check mode, proceed to Step 8. Otherwise, type
“
CM
” into the text box. Click “Send”.
4.
The tool
should display “CALIBRATE WRENCH”, “NO
“, “YES
”.
5.
Press
to enter calibration mode.
6.
Enter the CAL PASSWORD using
or
to change the value of the digit. Press
to move to the next digit. The default
password is “3112”. After all four digits have been entered, press
to submit the password.
7.
Press
until the torque check screen appears.
8.
The tool is in torque track mode to allow readings to be verified.
9.
Press
to advance to the angle check screen.
10.
The tool is in angle track mode. The tool displays the number of degrees the tool has been rotated. Press
to zero the
angle value on the tool at any time.
11.
Press
to exit angle check mode.
CALIBRATION COMPLETION
1.
The following procedure assumes the tool is still wirelessly connected to the radio dongle. Refer to the SR Connect
Setup section to reconnect if needed.
2.
Disable calibration mode by typing “
CF
” into the text box. Click “Send’. Click “Done”.
3.
Click “Forget tool”.
4.
Unscrew the battery holder and remove the battery.
5.
Screw the battery holder back onto the tool.
CLEANING
This tool should be cleaned with a soft cloth dampened with water. Do not immerse this tool in liquid or use any solvent
other than water to clean the tool.
CALIBRATION AND REPAIR
Factory calibration from our ISO 17025 Accredited Calibration Laboratory is available. Calibration and verification are
performed using the SR Connect PC application. Contact your SR distributor for details. Parts and factory repair are also
available.
ADDITIONAL INFORMATION
Additional information is available at
www.srtorque.com
. You can also call customer service at 1-847-455-8677 or send
e-mail to