Problem

Cause

Solution

Cooling liquid is not
circulated.

Cooling liquid filter is blocked.

Air is trapped in the pump.

Clean the filter – See the Maintenance and 
Servicing section for details.
Briefly remove the drain stopper.
until all trapped air has been removed.

Flow pressure is
reduced.

Tubing is trapped.
Pressure loss due to excessive 
tubing length.
The circulator is too low compared 
to the device it is cooling.

Check the tubing isn’t pinched, twisted or trapped
Try to reduce the overall length of the tubing.

Adjust the position of the recirculating cooler.

No refrigeration of
cooling medium.

Refrigeration unit is broken. 
The cooling medium isn’t sufficiently 
covering the cooling coil.
Overload of refrigeration unit 
(alarm led illuminates) 

Contact your local service representative.
Top up the cooling medium.

Lower ambient temperature below
35ºC, or raise set temperature.

Display does not 
illuminate when 
power is on.

No mains power.
Circuit breaker has tripped. 

Fuse has blown. 

Temperature controller is broken. 

Check mains power supply.
Unscrew maintenance panel and check 
switchable circuit breaker is in the on position. If 
problem persists there could be an electrical
leakage and you should contact your local service 
representative.
Follow procedure to change fuse in the 
Maintenance and services section
Contact local service representative

Cooling power is 
poor.

Refrigeration unit does not work. 
The ambient temperature exceeds 35ºC 
Exhaust is blocked. 

Refrigeration air filter is blocked 

Contact local service representative.
Try to lower the ambient temperature.
Check necessary clearance around the unit has 
been allowed.
Follow the procedure detailed in the Maintenance 
and Servicing section to clean the filter.

Water freezes in the 
circulation tank at 
+7ºC

Loss of flow pressure

Check flow pressure, see above
Use anti freeze for cooling medium.

Blinks on the display

Temperature sensor malfunction
Measured temperature is lower 
than the units lower limit (-50ºC)

Contact local service representative.

Blinks on the display, 
refrigeration stops

Temperature sensor short circuit
Measured temperature exceeds the 
upper limit of the unit (more than 
+40ºC)

The heating load exceeds capacity, reduce load.

Display is abnormal or 
unstable:

Electrical noise causes abnormal 
display on temperature controller
Temperature controller is reset by 
electrical noise 
Ambient temperature exceeds 35ºC 

Turn off mains power, wait 5 seconds then switch 
back on.
Move unit to a new location.

Lower ambient temperature below 35ºC.

- F F

E R

- - -

Instructions for use

Note: Alarm is cleared by turning off the power switch

Содержание SRC14

Страница 1: ...Recirculating Cooler SRC14 Instructions for use Version 2 0 ...

Страница 2: ...n provided may be impaired Electrical Installation THIS EQUIPMENT MUST BE EARTHED Before connection please ensure that the line supply corresponds to that shown on the rating plate This model requires a supply rated at 220 240V 50 Hz single phase Model Supply requirements Power SRC14 230V 50 Hz 1 1kW There is an IEC socket at the rear of the instrument for connection to the mains supply The unit i...

Страница 3: ...sive force connecting the hoses use universal clips or similar to secure the tubing if necessary Always consider the temperature and pressure requirements before selecting your tubing material Use tubing of the minimum length required excessively long tubing will reduce the efficiency of the unit Filling procedure When choosing your cooling medium you should take care that there is no risk of corr...

Страница 4: ... more than 5 seconds Pb will be displayed Using the up and down key to select the correct value as measured on the thermometer calibration can only be performed between 9 9ºC and 9 9ºC After setting the correct value press the set key for more than 5 seconds the display will return to the current measured temperature After operation Switch the circulation pump off then switch off the power switch ...

Страница 5: ...rformance and safety features designed into the instrument If in any doubt please contact the Technical Service Department of Bibby Scientific Ltd or the point of sale To open fusebox insert a small screwdriver into the slot and push downwards Drainage hose Fuse box AC outlet Drainage hose Fuse box AC outlet Bibby Scientific Ltd Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Fax ...

Страница 6: ...contact your local service representative Follow procedure to change fuse in the Maintenance and services section Contact local service representative Cooling power is poor Refrigeration unit does not work The ambient temperature exceeds 35ºC Exhaust is blocked Refrigeration air filter is blocked Contact local service representative Try to lower the ambient temperature Check necessary clearance ar...

Страница 7: ...fications We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that injury damage or loss might occur if equipment malfunctions due to radio frequency interference or for general advice before use contact the Technical Department of Bibby Scientific Ltd ...

Страница 8: ...225 9243 Fax 609 589 2571 www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketing bibby scientific it www bibby scientific it Bibby Scientific Asia Ltd Room 607 Yen Sheng Centre 64 Hoi Yuen Road Kwun Tong Kowloon Hong Kong Tel 852 3583 1581 Fax 852 3583 1580 e mail bibby bib...

Отзывы: