Operating manual Digital Transmitter DTM.OCS.S / DTM.OCS.S/N
Communication with the DTM.OCS.S transmitter
20
Doc-Nr. 10.00.0430
--------------------------------------------- data ---------------------------------------------
byte code:
50
20
31
30
2E
32
35
30
30
20
2D
50
55
interpretation:
‚P‘
‚ '
‚1‘
‚0‘
‚.‘
‚2'
‚5'
‚0'
‚0'
‚ '
‚-'
‚P'
‚U'
transmit sequence
--------------------------------------------- data ---------------------------------------------
byte code:
6D
48
32
4F
20
2D
54
20
32
37
2E
32
20
interpretation:
‚m‘ ‚H'
‚2‘
‚O‘
‚ ‘
‚-'
‚T'
‚ '
‚2'
‚7'
‚.'
‚2'
‚ '
transmit sequence
----------------------------------- data -----------------------------------
----- crc -----
byte code:
2D
54
55
20
C2
B0
43
20
4F
4B
3B
40
39
interpretation:
‚-‘
‚T'
‚U‘
‚ ‘
‚C‘
‚ ‘
‚O‘ ‚K‘
‚;‘
CRC
low
CRC
high
transmit sequence
Please note that UTF-8 encoding is used for the text strings. In the
example above this applies to the string
“°C”, resulting in the byte
code 0xC2B043 (marked cyan).
end of sequence