32
2040-309-002 REV B
i
BED Awareness System Error Codes
Error Handling
• The error handling system provides a mechanism for capturing errors known to the product and providing feedback
to the operator through logged errors or message errors.
• All Logged errors generate a generic blinking error message which will be displayed on the status screen in the
lower area: “Error XXX Call Service”, where XXX is the error code.
• Errors will only be logged if the error log does not already contain the error.
• In the event multiple errors are displayed on the status screen, the error code is used to identify the priority. The
lowest code has the highest priority.
• All logged errors and message errors are persistent until the power is cycled, or the condition is fixed.
• The details for each logged error and error message are detailed in the table below.
Logged Errors
Error
Code
Name
Description
Type
Affected Systems
100
Not Used
Highest Priority Errors
102
StuckSafetyRelay
This error is checked while the product is
“On”. If the safety relay signal is “On” and
there is no button press on the product then
set this error after 1 minute.” Note: This should
not be checked when in Auto-Cycle Mode.
Logged
Motion
103
InputTimer
If any footboard motion or side rail/pendant
input is sensed to be “On” continuously for a
total of 5 minutes then provide an exception
message. Note: This should not be checked
when in Auto-Cycle Mode.
Logged
Motion
200
Not Used
Scale System
201
HL-Load Cell
Patient Head Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
202
HR-Load Cell
Patient Head Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
203
FL-Load Cell
Patient Foot Left Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
204
FR-Load Cell
Patient Foot Right Load Cell. Under/Over
range.
Logged
Scale, Bed Exit,
i
BED Awareness
206
CS5533Fatal
Non-Recoverable
Communication
Error.
CS5533 Fatal occurs when the CPU fails
to communicate to the External ADC for the
Load cells and cannot recover.
Logged
Scale, Bed Exit,
i
BED Awareness
207
ZeroFail
Zero Fail error occurs when the system can-
not provide a stable zero value for three con-
secutive attempts at zeroing the product.
Logged
Scale, Bed Exit,
i
BED Awareness
Содержание Zoom 2040
Страница 2: ......
Страница 38: ...38 2040 309 002 REV B www stryker com Notes Return To Table of Contents ...
Страница 85: ...www stryker com 2040 309 002 REV B 85 Base Assembly and Options HEAD END Return To Table of Contents ...
Страница 86: ...86 2040 309 002 REV B www stryker com Base Assembly and Options Return To Table of Contents ...
Страница 102: ...102 2040 309 002 REV B www stryker com ZOOM Base Assembly HEAD END Return To Table of Contents ...
Страница 104: ...104 2040 309 002 REV B www stryker com ZOOM Base Assembly HEAD END HEAD END Return To Table of Contents ...
Страница 106: ...106 2040 309 002 REV B www stryker com ZOOM Base Assembly FOOT END HEAD END Return To Table of Contents ...
Страница 107: ...www stryker com 2040 309 002 REV B 107 ZOOM Base Assembly Return To Table of Contents ...
Страница 108: ...108 2040 309 002 REV B www stryker com ZOOM Base Assembly Return To Table of Contents ...
Страница 117: ...www stryker com 2040 309 002 REV B 117 Litter Assembly Return To Table of Contents For Reference Only 2032 231 010 ...
Страница 118: ...118 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Страница 119: ...www stryker com 2040 309 002 REV B 119 Litter Assembly Return To Table of Contents ...
Страница 120: ...120 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Страница 121: ...www stryker com 2040 309 002 REV B 121 Litter Assembly Return To Table of Contents ...
Страница 122: ...122 2040 309 002 REV B www stryker com Litter Assembly CPR RELEASE Return To Table of Contents ...
Страница 126: ...126 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents 2040 Head Wall View ...
Страница 127: ...www stryker com 2040 309 002 REV B 127 Litter Assembly Return To Table of Contents ...
Страница 128: ...128 2040 309 002 REV B www stryker com Litter Assembly Return To Table of Contents ...
Страница 146: ...146 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents For Reference Only 2040 031 012 ...
Страница 147: ...www stryker com 2040 309 002 REV B 147 ZOOM Litter Assembly DISPLAY HOUSING ASSEMBLY Return To Table of Contents ...
Страница 148: ...148 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Страница 149: ...www stryker com 2040 309 002 REV B 149 ZOOM Litter Assembly Return To Table of Contents ...
Страница 150: ...150 2040 309 002 REV B www stryker com ZOOM Litter Assembly Return To Table of Contents ...
Страница 151: ...www stryker com 2040 309 002 REV B 151 ZOOM Litter Assembly Return To Table of Contents ...
Страница 156: ...156 2040 309 002 REV B www stryker com Optional 110V Outlet Assembly Return To Table of Contents ...
Страница 158: ...158 2040 309 002 REV B www stryker com 110V Box Assembly Return To Table of Contents 2035 031 220 ...
Страница 160: ...160 2040 309 002 REV B www stryker com Head End Siderail Assembly Return To Table of Contents ...
Страница 161: ...www stryker com 2040 309 002 REV B 161 Head End Siderail Assembly CARRIER ARM ASSEMBLY Return To Table of Contents ...
Страница 162: ...162 2040 309 002 REV B www stryker com Head End Siderail Assembly CARRIER ARM ASSEMBLY Return To Table of Contents ...
Страница 176: ...176 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...
Страница 178: ...178 2040 309 002 REV B www stryker com Foot End Siderail Assembly Return To Table of Contents ...
Страница 217: ...www stryker com 2040 309 002 REV B 217 Optional Bed Extender Pad Return To Table of Contents 2025 040 010 ...