Instrument Bar
11
>
It is recommended that the anterior resection level be
checked to further confirm the correct size by sliding
a Bladerunner through the sizing guide’s size-specific
anterior slots and assessing the resection . If it appears
that there is a potential for notching, the next larger
size component will need to be chosen . Ensure the
femoral component chosen is compatible with the
size of the tibial component selected during tibial
preparation .
>
Once size confirmation is complete, attach the 1/8˝
Peg Drill to the Universal Driver and create fixation
pin-holes (for the 4:1 Cutting Block) through the
holes on the face of the Femoral Sizer marked “EPI” .
Femoral
Preparation
Figure 14
6541-4-003
Headless Pins – 3˝
6541-1-603
Femoral Sizer
6541-1-605
Femoral Stylus
6541-4-400
Bladerunner