28
Guidance and manufacturer's declaration - electromagnetic immunity
The TPS Console is intended for use in the electromagnetic environment specified below. The customer or the user of the TPS Console should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 6060
test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
T
)
for 0,5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
95% Reduction (10ms)
60% Reduction (100ms)
30% Reduction (500ms)
95% Reduction (5s)
3 A/m
@ 50Hz
CRT 1A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the TPS Console requires
continued operation during power mains
interruptions, it is recommended that the TPS
Console be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields should be at
levels characteristics of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Recommended separation distances between portable and mobile RF communications equipment and the TPS Console
The TPS Console is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the TPS Console can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the TPS Console as recommended below, according to the maximum output power of the
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
50 kHz to 80 MHz
d = [
—
]√P
0.12
0.37
1.17
3.70
11.70
80 MHz to 800 MHz
d = [
—
]√P
0.12
0.37
1.17
3.70
11.70
800 MHz to 2.5 GHz
d = [
—
]√P
0.23
0.74
2.33
7.37
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
Separation distance according to frequency of transmitter
m
3.5
V
3.5
E
7
E