Messages
Message
If you select OK to this question:
Delete all the surgeon's procedures?
The surgeon ID and all procedures will be deleted.
Change the name of all the surgeon's
Procedures listed under the surgeon will be moved under
procedures?
the current surgeon's name.
Identical record found.
Maximum number of records exceeded. Please delete inactive records.
Initializing console hardware. Please wait.
Handpiece has reached recommended service interval. Please return for service at earliest convenience.
Handpiece does not support the selected preference. Default handpiece setting will be used.
Handpiece temperature has exceeded its nominal operating range and may cause burning.
Handpiece temperature has exceeded its operating range. Allow to cool before restarting.
Procedure name unspecified.
Ensure handpiece speed does not exceed specified attachment limitations. Failure to do so may result in user and/or patient injury.
Cutter/bur is not compatible with the current handpiece.
Cutter/bur has exceeded its life. Please replace with a new cutter/bur.
Error Messages
Message
Action to Take
Error 00.
Console hardware fault detected.
Return console to Stryker for repair.
Error 002.
Console hardware fault detected.
Turn unit off and on again. If problem persists, return
Contact your Stryker service representative.
console to Stryker for repair.
Error 003.
Console hardware fault detected.
Turn console off and on again. If problem persists, return
Contact your Stryker service representative.
console to Stryker for repair.
Error 004.
Footswitch fault detected, right pedal will be disabled.
Unplug Footswitch from console and plug in again.
Contact your Stryker service representative.
If problem persists, return
footswitch to Stryker for repair.
Error 005.
Footswitch fault detected, left pedal will be disabled.
Unplug Footswitch from console and plug in again.
Contact your Stryker service representative.
If problem persists return
footswitch to Stryker for repair.
Error 006.
Footswitch unreadable.
Unplug Footswitch from console and
Contact your Stryker service representative.
plug in again. If problem persists,
return footswitch to Stryker for repair.
Error 007.
Handpiece fault detected, handpiece triggers will be
Unplug cord from handpiece and plug in again.
disabled. Contact your Stryker service representative.
If problem persists, return handpiece to Stryker
for repair.
Error 008.
Handpiece unreadable.
Unplug handpiece from console and plug in again.
Contact your Stryker service representative.
If problem persists, return handpiece Stryker for repair.
Error 009.
Handpiece requires additional console hardware.
Call your Stryker Instruments sales representative.
Contact your Stryker service representative.
Error 00
Handpiece requires additional console hardware.
Call your Stryker Instruments sales representative.
Contact your Stryker service representative.
Error 0
Handpiece requires additional console hardware.
Call your Stryker Instruments sales representative.
Contact your Stryker service representative.
Error 02
Cutting accessory requires additional console hardware.
See the instructions supplied with the cutting accessory.
The Total Performance System is not field repairable. In case of operating difficulties, Stryker products must be returned for maintenance or repair.