English
EN
www.stryker.com
101515 REV E
1-5
Introduction
Contraindications
Stryker promotes the assessment of each patient and appropriate usage by the operator. Contraindications for this
product are for patients with the conditions specified below.
Heating
contraindications include:
•
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
•
Application to a patient with an increased tendency to bleed (aggravates potential for hemorrhage)
•
Application to a body surface with possible malignancy (tissue metabolism is increased and therefore, the growth
potential of the malignant tissues)
•
Treatment of hematoma within the first 24 to 48 hours (potential for re-bleeding and hemorrhage). Recent sprain or
fracture (acute inflammatory response).
•
In combination with topical solutions whose toxicity may be affected by the application of heat
•
In combination with other heat sources
Cooling
contraindications include:
•
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
•
Application to body surface with known vascular impairment such as frostbite, arteriosclerosis or ischemia
•
Application to body surface in patients with hypersensitivity to cold, such as people with Raynaud’s phenomenon,
cold urticaria, cryoglobulinemia, and paraxysmal cold hemoglobinuria
•
Application to body surface in patients with impaired sensation
•
In combination with topical solutions whose toxicity may be affected by the application of cold
Specifications
Models
TP700 / TP700C
Temperature setpoints Fahrenheit
107° F, 100° F, 95° F, 50° F
Temperature setpoints Celsius
42° C, 38° C, 35° C, 10° C
Therapy duration
20 or 30 minutes or continuous cycle
Classification
Class I equipment with Type BF applied part suitable for continuous
operation. Not classified for protection against ingress of liquid. Not
classified for use in the presence of flammable anesthetics.
Size
11.5 in. x 8 in. x 8 in.
29.2 cm x 20.3 cm x 20.3 cm
Weight
Reservoir empty
6.5 lb
2.9 kg
Reservoir filled with water to the
heating level
9 lb
4.0 kg
Maximum reservoir capacity
93 oz
2.75 l
Flow rate, minimum with pad attached
9 gph
34 lph
Average temperature accuracy
±2° F at 107° F
±1° C to 42° C
Maximum contact surface temperature
107° F
42° C
High limit safety temperature
110° F to 120° F
43.3° C to 49° C
Power cord
International (harmonized) 3 wire cord set
Dual hose
10 ft
305 cm
Current leakage, maximum
100 µA
Ground resistance, maximum
0.5 mΩ
Electrical requirements
Voltage
120 VAC ±10%
Frequency:
60 Hz
Current in amperes
3.1
Содержание T/Pump Series
Страница 2: ...sample text ...
Страница 4: ...sample text ...
Страница 12: ...English EN 1 8 101515 REV E www stryker com Introduction Serial number location ...