73
Maintenance
Periodic Maintenance Schedule
WARNING
To ensure proper operation of the SwitchPoint Infinity® 2 you should perform the
following procedures periodically:
Activity:
Check leakage current to <500μA, (300μA in North America) ground impedance to <0.1 ohms, and power
consumption less than or equal to rated power.
When:
Every 12 months.
Tools and Equipment Required:
True RMS digital multimeter and safety analyzer.
Fuse Replacement Instructions
1. Unplug the SwitchPoint Infinity® 2 from the wall socket and remove the power cord from the unit.
2. Unlatch the fuse holder above the AC inlet and remove. You may need to press the tab on the fuse holder
with a slender screwdriver to release the latch.
3. Replace the fuse with the same value and rating.
WARNING
To avoid the risk of fire, replace only with a fuse of the value specified on the fuse label on
the back of the unit.
4. Reinstall the fuse holder until the tab snaps into place.
Cleaning Instructions
1. No periodic cleaning is required.
2. Metal surfaces requiring cleaning should be wiped down using 100% isopropyl alcohol.
3. To clean the touch panel display or the preview monitor use a soft cloth.
Caution
Do not use isopropyl alcohol or liquids of any kind on the touch panel display or the preview monitor. Do
not allow liquids to enter the case.
Disposition of the Product
The device must be disposed of according to local laws and hospital practices.
Removability of Batteries
The SwitchPoint Infinity® 2 consists of button cell batteries embedded into various boards in the system. These
button cells are not intended to be removed by the user or hospital. It is recommended that the SwitchPoint
Infinity® 2 be sent to the waste treatment facility. Refer to the WEEE Passports for the SwitchPoint Infinity® 2 for
recycling instructions.
The Batteries Directive 2006/66/EC introduces new requirements from September 2008 on removability of
batteries from waste equipment in EU Member States. To comply with this Directive, this SwitchPoint Infinity® 2
has been designed for safe removal of the batteries at end-of-life by a waste treatment facility. Infected units
should be decontaminated before they are sent for recycling.
StockCheck.com
Содержание SwitchPoint Infinity 2
Страница 35: ...33 S t o c k C h e c k c o m ...
Страница 61: ...59 S t o c k C h e c k c o m ...
Страница 80: ...S t o c k C h e c k c o m ...
Страница 81: ...S t o c k C h e c k c o m ...