
63
21.5 Troubleshooting High Resolution Switching
The SwitchPoint Infinity® 2 Control System can switch the input selection of the Stryker flat panel displays using
serial commands from any of the 16 RS-232 ports.
21.5.1 Flat Panel Not Switching Properly Between Video and High Resolution
•
Check that the DB-9 connector is connected securely to flat panel with serial input.
21.5.2 Image Not Properly Sized On Flat Panel
•
Check the resolution of the image you are sending. The maximum that a StrykerVision flat panel can
currently display is 1600X1200. The monitor will internally scale anything below that properly.
21.5.3 Wrong Image Being Displayed
•
Check that the flat panel has switched input mode. It can be manually switched if necessary.
StockCheck.com
Содержание SwitchPoint Infinity 2
Страница 35: ...33 S t o c k C h e c k c o m ...
Страница 61: ...59 S t o c k C h e c k c o m ...
Страница 80: ...S t o c k C h e c k c o m ...
Страница 81: ...S t o c k C h e c k c o m ...