SPY Portable Handheld Imaging System
Operator’s Manual
31
WARNING:
Proper sterile technique must be used when preparing the SPY-PHI imager
for surgical procedures. The following components are not sterile and should
not be handled by a sterile operator:
➢
Video Processor / Illuminator (VPI)
➢
Undraped SPY-PHI imager and cables.
WARNING:
Use only the approved drape model. Other drapes may result in poor imaging
performance.
WARNING:
The outside of the pouch containing the drape is NOT sterile.
WARNING:
Do not use the drape if the seals on the pouch appear compromised.
WARNING:
The drape is supplied sterile (sterilized using ethylene oxide) and must be
handled aseptically to maintain sterility.
WARNING:
The drape is supplied for single-patient use only. DO NOT RE-STERILIZE. DO
NOT RE-USE.
WARNING:
If the drape becomes compromised during application to the SPY-PHI imager
or during the procedure, remove the imager from the sterile field, remove the
compromised drape, and replace with a new drape as per the draping
instructions provided in this Operator's Manual.
CAUTION:
The drape is not intended to contact the patient. Do not contact the patient with the
draped SPY-PHI
device
.
Содержание SPY PHI HH9000
Страница 1: ......
Страница 6: ...iv 4 0003045 Rev G This page intentionally left blank...
Страница 8: ...2 4 0003045 Rev G This page intentionally left blank...
Страница 36: ...30 4 0003045 Rev G This page intentionally left blank...
Страница 58: ...52 4 0003045 Rev G This page intentionally left blank...
Страница 102: ...96 4 0003045 Rev G This page intentionally left blank...
Страница 106: ...100 4 0003045 Rev G This page intentionally left blank...
Страница 110: ...104 4 0003045 Rev G This page intentionally left blank...
Страница 111: ......
Страница 112: ......