![Stryker Prime TC Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/stryker/prime-tc/prime-tc_maintenance-manual_1386000015.webp)
P
Prro
occe
ed
du
urre
e::
1. Apply the brake.
2. Using a 3/16” hex wrench, remove the four flat head cap screws (A) and adaptors (B) to remove the seat back (C) from
the chair frame. Discard the screws, adaptors, and seat back (Figure 4).
A
B
A
B
D
Torque item A
to 70 ± 10 in-lb
Torque item A
to 70 ± 10 in-lb
C
F
Fiig
gu
urre
e 4
4 –– S
Se
ea
att b
ba
acckk rre
ep
plla
acce
em
me
en
ntt
3. Using a 3/16” hex wrench, install the four supplied flat head cap screws (A) and adaptors (B) to secure the supplied seat
back (C) to the chair frame. Using a 3/8” drive torque wrench, torque the screws to 70 ± 10 in-lb.
4. Apply the supplied Stryker label (D) in the appropriate color.
5. Verify proper operation before you return the product to service.
C
Ch
ha
arrtt h
ho
olld
de
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
5/32” hex wrench
•
3/8” drive torque wrench (in-lb)
P
Prro
occe
ed
du
urre
e::
1. Apply the brake.
2. Remove the seat back (see
3. Using a 5/32” hex wrench, remove the two button head cap screws (A) that secure the bottom of the chart holder to the
chair frame (Figure 5). Save the screws. When reinstalling the screws, use a 3/8” drive torque wrench to torque the
screws to 25 ± 5 in-lb.
1460-009-002 Rev B.0
11
EN
Содержание Prime TC
Страница 2: ......
Страница 4: ......
Страница 38: ...Torque item B to 70 10 in lb Torque item B to 70 10 in lb EN 34 1460 009 002 Rev B 0 ...