![Stryker Prime TC Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/stryker/prime-tc/prime-tc_maintenance-manual_1386000014.webp)
S
Se
errvviicce
e
S
Se
ea
att b
ba
acckk rre
em
mo
ovva
all
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
•
3/8” drive torque wrench (in-lb)
P
Prro
occe
ed
du
urre
e::
1. Apply the brake.
2. Using a 3/16” hex wrench, remove the four flat head cap screws (A) and adaptors (B) to remove the seat back (C) from
the chair frame. Save the screws, adaptors, and seat back (Figure 3).
A
B
A
B
Torque item A
to 70 ± 10 in-lb
Torque item A
to 70 ± 10 in-lb
C
F
Fiig
gu
urre
e 3
3 –– S
Se
ea
att b
ba
acckk rre
em
mo
ovva
all
3. When reinstalling the seat back, use a 3/16” hex wrench to reinstall the four flat head cap screws (A) and adaptors (B)
(removed in step 2) to secure the seat back (C) to the chair frame. Using a 3/8” drive torque wrench, torque the screws
to 70 ± 10 in-lb.
4. Verify proper operation before you return the product to service.
S
Se
ea
att b
ba
acckk rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
•
3/8” drive torque wrench (in-lb)
EN
10
1460-009-002 Rev B.0
Содержание Prime TC
Страница 2: ......
Страница 4: ......
Страница 38: ...Torque item B to 70 10 in lb Torque item B to 70 10 in lb EN 34 1460 009 002 Rev B 0 ...
Страница 63: ......
Страница 64: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 1460 009 002 Rev B 0 WCR AA 5 2019 09 ...