4405-210-700 Rev-AA
EN
www.stryker.com
7
Troubleshooting
WARNING:
DO NOT disassemble or service this equipment, unless otherwise specified.
NOTE:
For service, contact your Stryker sales representative or call Stryker customer service. Outside the US,
contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
ACTION
The handpiece operates but the
accessory does not move.
The attachment is not fully installed
in the handpiece.
Remove and install the attachment.
Make sure the attachment is
securely locked in the handpiece.
The attachment will not fit or cannot
be secured in the handpiece.
The attachment and/or the distal
end of the handpiece contains
debris.
Clean the attachment and/or
handpiece. See the care instructions
manual supplied with the equipment.
The attachment and/or handpiece is
damaged.
Return the attachment and/or
handpiece to Stryker for repair.
The attachment and/or cutting
accessory is wobbling in the
handpiece.
The accessory is the wrong size for
the attachment.
Install a different accessory or
attachment as required.
The accessory extends too far from
the distal end of the attachment
or is not properly centered in the
attachment.
Remove and properly install the
accessory.
The accessory is damaged.
Inspect the accessory for damage
and replace the accessory as
required.
The attachment and/or handpiece is
damaged.
Return the attachment and/or
handpiece to Stryker for repair.