Press (–) Button
Pressure Switch
Open?
Hold B-Valve open until
button released or pressure
switch closes
Open B-Valve
Open B-Valve and Ramp in
Motor
Ramp out motor and stop,
Close B-Valve
Bottom Hall Effects
Sensor Detected?
Run motor at max speed
Pressure Switch
Open?
Yes
No
No
No
Yes
Yes
L
Liitttte
err d
drriiffttss ((w
wiitth
ho
ou
utt p
pa
attiie
en
ntt w
we
eiig
gh
htt))
1. Check the manual release cable adjustment.
2. Change the locking manual valve.
3. Change “B” valve.
EN
24
6506-109-002 Rev G.1
Содержание Power-PRO XT
Страница 2: ......
Страница 6: ......
Страница 57: ...6506 109 002 Rev G 1 51 EN ...
Страница 58: ...EN 52 6506 109 002 Rev G 1 ...
Страница 59: ...6506 109 002 Rev G 1 53 EN ...
Страница 62: ...B Ba as se e a as ss se em mb blly y 6506 001 012 Rev AA Reference only EN 56 6506 109 002 Rev G 1 ...
Страница 91: ...P Po ow we er r a as ss se em mb blly y 6500 002 031 Rev D Reference only 6506 109 002 Rev G 1 85 EN ...
Страница 94: ...F Fo oo ot t e en nd d a as ss se em mb blly y 6500 102 015 Rev AA Reference only EN 88 6506 109 002 Rev G 1 ...
Страница 103: ...6506 109 002 Rev G 1 97 EN ...
Страница 122: ...EN 116 6506 109 002 Rev G 1 ...
Страница 145: ...P Piillllo ow w 6 61 10 00 0 0 04 41 1 0 03 30 0 Rev E Reference only 6506 109 002 Rev G 1 139 EN ...
Страница 147: ...W Wh he ee ell f fr ro on nt t 6 65 50 00 0 0 00 01 1 0 08 86 6 Rev A Reference only 6506 109 002 Rev G 1 141 EN ...
Страница 166: ...T Tr ra an ns sf fe er r f flla at t 6005 001 001 Rev E Reference only EN 160 6506 109 002 Rev G 1 ...
Страница 185: ......
Страница 186: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6506 109 002 Rev G 1 WCR AB 11 2019 07 ...