Remanufactured LigaSure Exact Dissector, Without Nano-coating 

Page 6 of 9 

 

 

4.

 

To seal adjacent tissue, overlap the edge of the existing seal. The second seal should be distal to the first seal to 

increase seal margin. 

 

Footswitch Activation

 

A footswitch can be used instead of the hand-activation button (3). Ensure that the footswitch is connected to the footswitch 

receptacle that corresponds to the instrument receptacle in use. 

 

Warning

 

Activating energy delivery with a footswitch when the hand-activation button is not in the second-click position may result in 

improper sealing and increase thermal spread to tissue outside the surgical site. The second click of the hand-activation button 

indicates proper pressure is being applied to the tissue. 

 

1.

 

Connect the footswitch to the generator as described in the generator user’s guide. 

2.

 

Ensure hand-activation is disabled on the appropriate port. Refer to the generator user’s guide if needed. 

3.

 

Grasp the intended vessel and/or tissue in the center of the jaws. 

4.

 

Press and hold ring handles to the second click to ensure appropriate pressure on the grasped tissue. 

5.

 

Press and hold the footswitch pedal to activate energy. 

 

A continuous tone sounds to indicate the activation of RF energy. When the activation cycle is complete, a two-pulsed Seal-

Cycle-Complete tone sounds and RF output ceases. 

 

Notice

 

The surgeon may inspect the seal before cutting the vessel or tissue. After inspecting the seal, the surgeon should create a 

second seal adjacent to the first seal before cutting, as described below. 

 

A tone with multiple pulses indicates that the seal cycle was not completed. Refer to the Troubleshooting section for 

possible causes and corrective actions. Do not cut tissue until you have verified that there is an adequate seal.

 

 

6.

 

To seal adjacent tissue, overlap the edge of the existing seal. The second seal should be distal to the first seal to 

increase seal margin. 

 

Cutting Tissue

 

Warning

 

Energy-based devices, such as ES pencils or ultrasonic scalpels that are associated with thermal spread, should not be used ot 

transect seals. 

 

Notice

 

Failure to maintain steady pressure on the ring handles while cutting could result in inadvertent reactivation of energy. 

 

1.

 

To divide tissue, maintain steady pressure on the ring handles and pull the cutting trigger (4) until a hard stop is 

reached. Then release the cutting trigger to allow the cutting blade to retract. 

2.

 

Open the jaws to release tissue. 

 

Cleaning the Instrument During Use

 

Warning

 

Inspect the instrument jaws prior to cleaning to ensure the cutting blade is not deployed. 

 

Do not activate the instrument or cutting trigger while cleaning the jaws. Injury to operating room personnel may result. 

 

Wipe jaw surfaces and edges with a wet gauze pad as needed. 

 

Notice

 

Do not attempt to clean the instrument jaws by activating the instrument on wet gauze. Product damage may occur. 

 

Remove any embedded tissue from blade track and jaw hinge area. 

 

Do not clean the instrument jaws with a scratch pad or other abrasives. 

 

Содержание LF2019

Страница 1: ...entified ISO 15223 1 2016 2492 Batch code Indicates the manufacturer s batch code so that the batch or lot can be identified ISO 15223 1 2016 1641 Consult instructions for use Indicates the need for the user to consult the instructions for use ISO 15223 1 2016 1051 Do not re use Indicates a medical device that is intended for one use or for use on a single patient during a single procedure ISO 152...

Страница 2: ... and division of vessels lymphatics and tissue bundles 2 3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures Do not use the LigaSure system for these procedures Contraindications for Use The LigaSure system has not been shown to...

Страница 3: ...ach side arrow shows direction for cutting Cable cord Connector purple and white Precaution The nano coating featured on the original device is not present on this product Adjunctive use of nerve monitoring device is recommended during nerve sparing procedures such as thyroidectomies radical neck dissection and parotidectomies Secondary hemorrhaging after tonsillectomy is a potentially serious adv...

Страница 4: ...nerator Refer to the generator user s guide for additional set up information During Surgery The LigaSure Sealer Divider can be used during surgery both to manipulate and dissect tissue and to seal and cut vessels and tissue bundles Instructs for use of the instrument during a procedure are provided in this section Warning Avoid placing fingers between the jaws or ring handles Injury to the use ma...

Страница 5: ...the system only when the instrument is in direct contact with the target tissue to lessen the possibility of unintended burns Do not activate the instrument while instrument jaws are in contact with or in close proximity to other instruments including metal cannulas as localized burns to the patient or physician may occur Eliminate tension on the tissue when sealing and cutting to ensure proper fu...

Страница 6: ...ssel or tissue After inspecting the seal the surgeon should create a second seal adjacent to the first seal before cutting as described below A tone with multiple pulses indicates that the seal cycle was not completed Refer to the Troubleshooting section for possible causes and corrective actions Do not cut tissue until you have verified that there is an adequate seal 6 To seal adjacent tissue ove...

Страница 7: ... The user is grasping thin tissue or not enough tissue open the jaws and confirm that a sufficient amount of tissue is inside the jaws If necessary increase the thickness of the tissue that is grasped and reactivate the seal cycle Too much tissue between the jaws The user is grasping too much tissue open the jaws reduce the amount of tissue tat is grasped and reactivate the seal cycle Activating o...

Страница 8: ...of one year from the date of purchase General Warranty Terms Applicable to All Products TO THE FULLEST EXTENT PERMITTED BY LAW THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY WARRANTY APPLICABLE TO THE PRODUCTS AND IS EXPRESSLY IN LIEU OF ANY OTHER WARRANTY BY STRYKER EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IN ...

Страница 9: ...he initial warranty period has expired by the time the product is repaired or replaced for thirty 30 days after delivery of the repaired or replaced product When a product or component is replaced the item provided in replacement will be the customer s property and the replaced item will be Stryker s property If a refund is provided by Stryker the product for which the refund is provided must be r...

Отзывы: