Remanufactured LigaSure Exact Dissector, Without Nano-coating
Page 3 of 9
The safe and effect use of RF energy depends on many factors solely under the control of the operator. There is no substitute
for properly trained and vigilant personnel. It is important that the operating instructions supplied with this or any other
medical equipment be read, understood, and followed
Precaution
Use caution during surgical cases in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurysmal
vessels, etc.). For best results, apply the seal to unaffected vasculature.
①
Jaws (gray)
②
Ring handles (gray)
③
Hand-activation button (purple)
④
Cutting triggers (gray), one on each side (arrow shows direction for cutting)
⑤
Cable/cord
⑥
Connector (purple and white)
Precaution
The nano-coating featured on the original device is not present on this product.
Adjunctive use of nerve-monitoring device is recommended during nerve-sparing procedures such as thyroidectomies, radical
neck dissection and parotidectomies.
Secondary hemorrhaging after tonsillectomy is a potentially serious adverse event. According to some reports, electrosurgical
devices may be associated with a small increase in the incidence of secondary post-tonsillectomy hemorrhaging.
Set-up
Warning
Electric Shock Hazard
– Do not connect wet accessories to the LigaSure system.
Position instrument cords to avoid contact with the patient or other cords. Do not wrap cords around metal objects. This may
induce currents that could lead to shocks, fires, or injury to the patient or surgical team.