NNoottee
• When handling the pod sensor hoses, do not bend or kink the hoses.
• Pay attention to the sensor hose position and insertion color and number labels.
• During reinstall, insert the hose and once you hit a stop, continue to push until you reach a hard stop.
Y
FFiigguurree 77 –– PPoodd sseennssoorr hhoosseess aanndd ppuurrssee lloocckk wwiirree ttiiee
9. Remove sensor hoses from the purse lock wire ties (M) (Figure 7).
10.Reach under the pod assembly (C) from the patients right side near the middle to access the turn bladder and
disconnect the turn bladder quick disconnect. Repeat on the patient left side (Figure 8).
2972-009-002 Rev AA.0
29
EN
Содержание Isolibrium 2972
Страница 2: ......
Страница 22: ...I It te em m C Co ollo or r 1 Black 2 Red 4 Green 6 open EN 18 2972 009 002 Rev AA 0 ...
Страница 54: ...EN 50 2972 009 002 Rev AA 0 ...
Страница 55: ...2972 009 002 Rev AA 0 51 EN ...
Страница 56: ...EN 52 2972 009 002 Rev AA 0 ...
Страница 57: ...2972 009 002 Rev AA 0 53 EN ...
Страница 58: ...EN 54 2972 009 002 Rev AA 0 ...
Страница 62: ...Torque item B to 22 4 in lb Torque item B to 22 4 in lb EN 58 2972 009 002 Rev AA 0 ...
Страница 63: ...Torque item A to 22 4 in lb Tighten item BG force 25 40 lb 2972 009 002 Rev AA 0 59 EN ...
Страница 65: ...Torque item C to 8 2 in lb 2972 009 002 Rev AA 0 61 EN ...
Страница 66: ...Torque item B to 22 4 in lb EN 62 2972 009 002 Rev AA 0 ...
Страница 75: ...2 29 97 71 1 0 02 22 2 0 00 08 8 Rev AB Reference only AB 2971 021 033 1 2972 009 002 Rev AA 0 71 EN ...
Страница 76: ...2 29 97 71 1 0 02 22 2 0 00 08 8 Rev AB Reference only AJ 297300220014 1 EN 72 2972 009 002 Rev AA 0 ...
Страница 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2972 009 002 Rev AA 0 WCR AA 5 2021 07 ...