5. Using a #2 Phillips screwdriver, remove the screw (B) that secures the hook assembly to the back pump housing. Save
the screw.
6. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
7. Rotate the pump assembly to the right, so the front of the pump is on your left.
8. Separate the front and back pump housing. Place the back of the pump down on the work surface.
9. Using a #2 Phillips screwdriver, remove the eight screws that secure the mounting plate to the back pump housing.
Save the screws.
10. Using a #2 Phillips screwdriver, remove the one screw that secures the p-clamp and compressor wires to the mounting
plate. Save the screw.
11. Tip the mounting plate in toward the front pump housing to access the transport handle screws.
12. Using a #2 Phillips screwdriver, remove the six screws (E) that secure the hook assembly (D) to the back pump housing.
Save the screws.
13. Remove and discard the hook assembly.
14. Reverse steps to reinstall.
15. Plug the pump cord into an accessible outlet.
16. Run the pump diagnostic test. All tests must pass.
17. Verify proper operation before you return the product to service.