H
Ho
osse
e sslle
ee
evve
e rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
None
P
Prro
occe
ed
du
urre
e::
1. Disconnect the CPR connector and the AST cable from the pump.
2. Raise the bed height to the highest position.
3. Place the litter surface in a flat, horizontal position.
4. Unbutton the hose sleeve from the bottom cover.
5. Slide the hose sleeve (A) over the CPR connector and remove (Figure 11). Discard hose sleeve.
F
Fiig
gu
urre
e 1
11
1 –– H
Ho
osse
e sslle
ee
evve
e
6. Slide the hose sleeve over the CPR connector.
7. Button the hose sleeve to the bottom cover.
8. Verify proper operation before you return the product to service.
C
CP
PR
R cco
on
nn
ne
ecctto
orr a
an
nd
d h
ho
osse
e rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Diagonal pliers
•
Rubbing alcohol
P
Prro
occe
ed
du
urre
e::
1. Raise the bed height to the highest position.
2. Place the litter surface in a flat, horizontal position.
3. Unzip and remove the support surface top cover.
4. Unclip the foot end center belt retainers from the bottom cover.
5. Remove the CPR connector to allow the support surface to deflate.
6. Unzip the fire barrier.
7. Move the fire barrier over the side bolsters and air cells toward the head end to access the end of the tubing.
8. Using diagonal pliers, cut the four nylon cable ties that secure the hoses in the support surface to the CPR connector.
9. Grasp the hose and connector on the support surface side, and pull each hose from the pump connector to remove.
N
No
otte
e
•
Each hose has a label near the end of the hose and indicates the function. Make sure that you note each hose
connection before removal.
•
Use rubbing alcohol to make the hose fit easier when you reinstall.
EN
18
2941-109-002 Rev AA.0