EN-34
Disposable
Attachments
RF probe is not
ready
Check the connection to the
console.
RF probe is expired
Replace probe.
RF probe
identification is
invalid
Replace probe.
RF probe
communication
error
• Check the connection to the
console.
• If necessary, replace probe.
Exceeded time
usage
Replace probe
RF power is too
high
• Check the probe for damage.
• If necessary, replace probe.
RF voltage is too
high
• Check the probe for damage.
• If necessary, replace probe.
RF current is too
high
• Check the probe for damage.
• If necessary, replace probe.
RF delivery
has exceeded
continuous limit
Clear error and continue
Low impedance
detected
• Check the probe for damage.
• If necessary, replace probe.
Footswitch
A wireless
footswitch is
detected
Disconnect the wired footswitch.
The footswitch icon
does not appear
• Ensure the unit is connected.
• Ensure that there is no damage to
the cable or connector.
Note:
If a disturbance occurs on the video monitor, the user should ensure that
the probe cable is not near any other instrument cables.