2012/12 Rev C.0

100259 Rev C

Page 1 of 2

Care•Free Companion

®

 Pump  

 CF300

FUNCTIONAL CHECK INSTRUCTIONS

Read and understand all instructions and the following safety precautions before beginning any work.

SAFETY PRECAUTIONS

 DANGER

•  Electrical shock hazard. Do not remove cover; there are no serviceable parts inside.
•  Explosive hazard. Do not use in the presence of flammable anesthetics.

TOOLS REQUIRED

Product

Tool

Description

CF300

Tester

P/N 12157

Voltage: 120V

Functional Check Instructions

P/N 100259

Frequency: 60Hz

Manometer

0-50 mm Hg or greater (standard mercury, or digital)

Tubing

3/16” ID

GENERAL DESCRIPTION

The 

Functional Check

 Instructions provide a 

quantitative

 means of determining whether a Care•Free Companion 

Pump can properly inflate its cushion. An alternate, 

qualitative 

means to check pump and cushion performance is to 

follow the HAND CHECK described in the 

Care•Free Companion Pump 

Instructions for Use.

The 

Functional Check

 

Instructions

 require the use of a Tester tool that simulates the cushion that is normally used with 

these pumps. If the pump does not meet the outlined requirement, it should be replaced. Repair is not available for 
these pumps.

PHYSICAL INSPECTION

1.  Unplug the power cord.

2.  Check tubing and cap for visible signs of damage (kinks, cuts, and cracks). 

 WARNING

•  Disinfect before use with each patient, utilizing standard hospital protocol and disinfectants. Failure to disinfect may 

cause cross-contamination and infection.

•  Do not autoclave. Autoclaving will damage the pump.

 CAUTION

•  The Functional Check is an optional test, however should you choose to perform the test it must be performed only 

by qualified technical personnel familiar with good practice for medical device repair.

•  Verify proper operation of the pump. Operation of the pump at less than specified flows and pressures may result 

in poor or reduced patient therapy.

Warning - Consult 
accompanying documentation

Manufacturer

Caution

Model

Shock Hazard

Do not use with Flammable Anesthetics