A
B
D
C
E
F
Fiig
gu
urre
e 1
1 –– B
Bu
um
mp
pe
errss
4. Using G
Go
oo
off O
Offff Household Heavy Duty Remover, a small slotted screwdriver, and towels, remove the adhesive from
the back cover.
5. Using alcohol prep wipes, clean the back cover bumper pad areas and current compliance label (E).
6. Install the new top bumper (A), bottom left bumper (B), and bottom right bumper (C) to the back cover (D).
7. Verify proper operation before you return the product to service.
F
Fo
oo
ott rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
6 mm socket
•
Ratchet
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Using two hands, rotate the pump assembly to the right so the front is on your left.
6. Separate the front and back pump housing. Place the back pump housing down on the work surface.
7. Using a ratchet and a 6 mm socket, remove the six nuts that secure the foot to the front pump housing. Save the nuts.
8. Remove and discard the foot.
9. Reverse the steps to reinstall (
Pump housing rubber foot assembly - 2874-007-022
10. Verify proper operation before you return the product to service.
2863-009-002 Rev AA.1