Maintenance Instructions
6
CLEANING INSTRUCTIONS
WARNING
Unplug the power cord from the wall socket before performing any maintenance on the light or personal injury
could result.
1.
Turn off power to the light and unplug the power cord from the wall socket.
2.
Use a damp cloth and mild cleaning solution to wipe soiled areas.
CAUTION
Do not submerge or autoclave the light assembly or damage to the unit could result.
BULB REPLACEMENT
WARNING
Unplug the power cord from the wall socket before performing any maintenance on the light or personal injury
could result.
1.
Turn off power to the light and unplug the power cord from the wall socket.
2.
Remove the three screws holding the lens and shroud. Remove the lens and shroud.
3.
Ensure the bulb is not hot and remove it from the light.
4.
Install a new bulb, using a clean cloth to handle the bulb. Do not touch the bulb with bare fingers, since
skin oils can cause premature bulb failure. Do not move the bulb from side to side while installing it; push
it straight in.
5.
Replace the shroud, lens, and the three screws to hold them in place.
6.
Plug the power cord into the wall socket and turn on the light.
Bulb reorder number: 88–1133–15–00