Stryker Adel Скачать руководство пользователя страница 3

Installation Instructions

3

INSTALLING THE INFANT WARMER CARE STATION LIGHT

Required Tools:

Drill with 3/32” Bit

Phillips Screwdriver

Standard Screwdriver

6mm Allen Wrench

Procedure:

1. If the Infant Warmer being worked on was ordered with a light, and has pilot holes already installed, omit

step two.

2. Facing the front of the Infant Warmer, the light must be mounted on the left side (opposite the control pan-

el).  Using the drill and the 3/32” bit, drill four pilot holes as shown in diagram B.

WARNING

Mount the light only in the specified location.

3. Install the light mount as shown in diagram D, using the four 1 1/4” screws supplied in the kit.  The light

pivot screw should face the front of the unit.

4. Push the light base stud into the mount and tighten the pivot screw, using either a standard screwdriver

or a 6mm Allen wrench.

5. Affix the warning labels to the plastic diffuser tray as shown in diagram A.

6. Using the dimensions in diagram C, drill two pilot holes for the cord cleat.  Install the cleat with the screws

supplied in the kit.

Содержание Adel

Страница 1: ...IMPORTANT File in your maintenance records Infant Warmer Care Station Cabinet Light OPERATIONS AND MAINTENANCE MANUAL For Parts or Technical Assistance 1 800 327 0770 ...

Страница 2: ...F THE BEAM MAY SHINE INTO THE INFANT S EYES Label 2 WARNING WHEN LIGHT IS IN USE THE INFANT SHOULD ALWAYS BE POSITIONED WITH HEAD AT THIS END OF BED See page 4 for installation instructions for these labels SPECIFICATIONS 120V 0 2A 60Hz 20VA 12V 1 7A Light Bulb 12v 20w Halogen WARNING CAUTION NOTE DEFINITION The words WARNING CAUTION and NOTE carry special meanings and should be carefully reviewed...

Страница 3: ...the drill and the 3 32 bit drill four pilot holes as shown in diagram B WARNING Mount the light only in the specified location 3 Install the light mount as shown in diagram D using the four 1 1 4 screws supplied in the kit The light pivot screw should face the front of the unit 4 Push the light base stud into the mount and tighten the pivot screw using either a standard screwdriver or a 6mm Allen ...

Страница 4: ...DIAGRAM B Label 1 WARNING WHEN LIGHT IS IN USE DO NOT POSITION THE LIGHT SO THE CENTER OF THE BEAM MAY SHINE INTO THE INFANT S EYES Label 2 WARNING WHEN LIGHT IS IN USE THE INFANT SHOULD ALWAYS BE POSITIONED WITH HEAD AT THIS END OF THE BED ...

Страница 5: ...Installation Instructions 5 DIAGRAM C DIAGRAM D ...

Страница 6: ...ore performing any maintenance on the light or personal injury could result 1 Turn off power to the light and unplug the power cord from the wall socket 2 Remove the three screws holding the lens and shroud Remove the lens and shroud 3 Ensure the bulb is not hot and remove it from the light 4 Install a new bulb using a clean cloth to handle the bulb Do not touch the bulb with bare fingers since sk...

Страница 7: ...AL DAMAGES ARISING FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT To Obtain Parts and Service Stryker Adel products are supported by a nationwide network of dedicated Stryker Adel Field Service Repre sentatives These representatives are factory trained available locally and carry a substantial spare parts inventory to minimize repair time Simply call your local representative ...

Страница 8: ...fteen 15 days of receipt of merchandise DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT Upon prompt notification Stryker Adel will file a freight claim with the appropriate carrier for damages incurred Claim will be limited in amount to the actual replacement cost In the event that this information is not received by Stryker Adel within th...

Страница 9: ...DH 5 96 1133 14 00 REV B 6300 Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 ...