Guidance and manufacturer's declaration - electromagnetic immunity
The Cast Cutter REF 941 is intended for use in the electromagnetic environment specified below. The customer or the user of
the Cast Cutter REF 941 should assure that it is used in such an environment.
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5%
U
T
(>95% dip in
U
T
)
for 0,5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 sec
3 A/m
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance level
±2, 4,6 kV contact
±2,4, 8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
95% Reduction (10ms)
60% Reduction (100ms)
30% Reduction (500ms)
95% Reduction (5s)
3 A/m
@ 50Hz
CRT 1A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Cast Cutter REF 941
requires continued operation during power
mains interruptions, it is recommended that
the Cast Cutter REF 941 be powered from
an uninterruptible power supply or a battery.
Power frequency magnetic fields should
be at levels characteristics of a typical
location in a typical commercial or hospital
environment.