9
NOTE:
Refer to
Replacement Parts List
for component
information.
1. Disconnect the Cast Cutter from the electrical supply.
2. Use a screwdriver to loosen and remove the brush
caps located on each side of the Cast Cutter. Remove
and discard the used springs and brushes.
3. IMPORTANT: Insert the new brushes oriented so that
the concave surface of the brush is positioned as
indicated in the illustration. If a brush is rotated, it will
not make proper contact.
4. Place the springs onto the brush and install the brush
caps, tighten brush caps to 4-8 in-lbs (.45-.90 N·m).
5. Verify that leakage current complies with appropriate
specification.
Periodic lubrication or calibration is not required. The necessary lubrication and calibration adjustments have been made in the
factory.
WARNING:
Do not use the Cast Cutter if leakage current exceeds appropriate specification (see
Specifications
). Return your
Cast Cutter to Stryker for service. (Outside US, return it to the nearest Stryker subsidiary).
Periodic Maintenance
Activity/Item
Source cord
3-way switch
RPM check
Interval
Annually
Annually
Annually
Tools and Equipment Required
None required
None required
Use RPM tester for correct speed
Instructions
Check for cracks. Return for repair if required.
Stop and start unit at least 5 times.
See
Specifications
.
Brush Replacement