Cleaning
Hand wash all surfaces of the stretcher with warm water and mild detergent. dry thoroughly. Do not steam clean,
pressure wash, hose off or ultrasonically clean the stretcher. Using these methods of cleaning is not recommended and
may void this product’s warranty.
Suggested cleaners for stretcher surfaces:
Quaternary Cleaners (active ingredient - ammonium chloride).
Phenolic Cleaners (active ingredient - o-phenylphenol).
Chlorinated Bleach Solution (5.25% - less than 1 part bleach to 100 parts water).
Avoid oversaturation and ensure the product does not stay wet longer than the chemical manufacturer’s guidelines for
proper disinfecting.
CaUTIOn
SOMe CleanIng prOdUCTS are COrrOSIve In naTUre and May CaUSe daMage TO THe prOdUCT If USed
IMprOperly
. If the products suggested above are used to clean Stryker patient care equipment, measures must be
taken to insure the stretcher is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning.
Failure to properly rinse and dry the stretcher will leave a corrosive residue on the surface of the stretcher, possibly
causing premature corrosion of critical components. Failure to follow the above directions when using these types of
cleaners may void this product’s warranty.
For mattress cleaning instructions, please see the tag on the mattress, or contact the mattress manufacturer.
Clean Velcro
®
after each use. Saturate Velcro
®
with disinfectant and allow disinfectant to evaporate. (Appropriate
disinfectant for nylon Velcro
®
should be determined by the hospital).
•
•
•