All fasteners secure.
Siderails move and latch properly.
Engage the brake pedal and push on the stretcher to ensure all casters lock securely.
All casters secure and swiveling properly.
Body restraints working properly.
I.V. pole intact and operating properly.
Oxygen bottle holder intact and operating properly.
Fowler operates and latches properly.
Trendelenburg/Reverse Trendelenburg operating properly.
No rips or cracks in mattress cover.
Ground chain intact.
No leaks at hydraulic connections.
Hydraulic jacks holding properly.
Hydraulic drop rate set properly.
Hydraulic oil level sufficient.
Lubricate where required.
Accessories and mounting hardware in good condition and working properly.
CHeCklIST
Stretcher Serial Number:
Completed by: _______________________________________
Date: _________________
note
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.