Cleaning
6
Hand wash all surfaces of the stretcher with warm water and mild detergent. Dry thoroughly. DO NOT
STEAM CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN. Using these methods
of cleaning is not recommended and may void this product’s warranty.
Clean Velcro AFTER EACH USE. Saturate Velcro with disinfectant and allow disinfectant to evaporate. (Ap-
propriate disinfectant for nylon Velcro should be determined by the hospital.)
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or
quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use be-
cause staining may result. The following products have been tested and have been found not to have a harm-
ful effect WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME
DISINFECTANT
TYPE
MANUFACTURER
*MANUFACTURER’S
RECOMMENDED
DILUTION
A33
Quaternary
Airwick (Professional Products Division)
2 ounces/gallon
A33 (dry)
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Beaucoup
Phenolic
Huntington Laboratories
1 ounce/gallon
Blue Chip
Quaternary
S.C. Johnson
2 ounces/gallon
Elimstaph
Quaternary
Walter G. Legge
1 ounce/gallon
Franklin Phenomysan F2500
Phenolic
Purex Corporation
1 1/4 ounce/gallon
Franklin Sentinel
Quaternary
Purex Corporation
2 ounces/gallon
Galahad
Phenolic
Puritan Churchill Chemical Company
1 ounce/gallon
Hi−Tor
Quaternary
Huntington Laboratories
1/2 ounce/gallon
LPH
Phenolic
Vestal Laboratories
1/2 ounce/gallon
Matar
Phenolic
Huntington Laboratories
1/2 ounce/gallon
Omega
Quaternary
Airwick (Professional Products Division)
1/2 ounce/gallon
Quanto
Quaternary
Huntington Laboratories
1 ounce/gallon
Sanikleen
Quaternary
West Chemical Products
2 ounces/ gallon
Sanimaster II
Quaternary
Service Master
1 ounce/gallon
Vesphene
Phenolic
Vestal Laboratories
1 1/4 ounce/ gallon
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So-
dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach
to 100 parts water are not considered mild detergents. These products are corrosive in nature and
may cause damage to your stretcher if used improperly. If these types of products are used to clean
Stryker patient handling equipment, measures must be taken to insure the stretchers are rinsed with clean
water and thoroughly dried following cleaning. Failure to properly rinse and dry the stretchers will leave a cor-
rosive residue on the surface of the stretcher, possibly causing premature corrosion of critical components.
NOTE
Failure to follow the above directions when using these types of cleaners may void this product’s warranty.
REMOVAL OF IODINE COMPOUNDS
This solution may be used to remove iodine stains from mattress cover surfaces.
1. Use a solution of 1−2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface.
2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.
Содержание 736
Страница 17: ...0853 006 130 Fifth Wheel Assembly 16 Return to Table of Contents ...
Страница 25: ...Side Control Brake Assembly 24 Assembly part number 0853 003 220 reference only Return to Table of Contents ...
Страница 26: ...Side Control Brake Assembly 25 See Detail A Detail A Typical Both Ends Return to Table of Contents ...
Страница 32: ...Dual Sided Hydraulics Assembly Fifth Wheel Base 31 See Detail A Detail A Return to Table of Contents ...
Страница 40: ...Notes 39 Return to Table of Contents ...
Страница 44: ...Litter Assembly 43 Assembly part number 0736 010 010 reference only Return to Table of Contents ...
Страница 45: ...Litter Assembly 44 Return to Table of Contents ...
Страница 46: ...Litter Assembly 45 Return to Table of Contents ...