![Stryker 7205-000-000 Скачать руководство пользователя страница 3](http://html.mh-extra.com/html/stryker/7205-000-000/7205-000-000_instructions-for-use-manual_1386105003.webp)
7205-001-700 Rev-AA
EN
www.stryker.com
3
• Upon initial receipt and before each use, operate
the equipment and inspect each component for
damage. DO NOT use any equipment if damage
is apparent. See the
Heavy Duty Care Instructions
manual for inspection criteria.
• DO NOT use this equipment in areas in which
flammable anesthetics or flammable agents are
mixed with air, oxygen, or nitrous oxide.
• Take special precautions regarding electromagnetic
compatibility (EMC) when using medical electrical
equipment. Place this equipment into service
according to the EMC information contained in this
manual. Portable and mobile radio frequency (RF)
equipment can affect the function of this equipment.
• DO NOT stack or place equipment adjacent to
the product. If such a configuration is necessary,
observe the configuration to ensure that
electromagnetic interference does not degrade
performance.
• DO NOT use the product in a magnetic resonance
imaging (MRI) environment. Using the product in an
MRI environment could affect the function of the
system.
Accessories
WARNINGS:
• Use only Stryker-approved system components and
accessories, unless otherwise specified. DO NOT
modify any system component or accessory.
• Using other electronic components and accessories
may result in increased electromagnetic emissions
or decreased electromagnetic immunity of the
system.
• DO NOT reuse, reprocess, or repackage a device
that is intended for single use only.
-
A single use device may not withstand
chemical, chemical vapor, or high temperature
sterilization reprocessing.
-
Design features may make cleaning difficult.
-
Reuse may create a contamination risk and
compromise structural integrity resulting in
operational failure.
-
Critical product information may be lost during
repackaging.
Failure to comply may lead to infection or cross
infection and result in patient and/or healthcare staff
injury.
NOTES:
• A variety of attachments are available for use with
this handpiece. Each attachment has a specialized
retainer for wires, pins, tools and/or cutting
accessories. See the
Heavy Duty Rotary Handpiece
Attachments Instructions For Use
manual for specific
attachment and accessory instructions.
• For a complete list of accessories, contact your
Stryker sales representative. Outside the US,
contact your nearest Stryker subsidiary.