Troubleshooting (continued)
PROBLEM
CAUSE
ACTION
The attachment will
not fit or cannot
be secured in the
handpiece.
The attachment and/or the distal end of
the handpiece contains debris.
Use a small brush with stiff, non-metallic
bristles to clean the attachment and/or the
distal end of the handpiece.
The attachment is damaged.
Return the attachment to Stryker for repair.
The handpiece is damaged.
Return the handpiece to Stryker for repair.
The attachment and/
or accessory wobbles
in the handpiece.
The accessory is damaged.
Inspect the accessory for damage and
replace the accessory as required.
The accessory is the wrong size for the
attachment.
Install a different accessory or attachment
as required.
The accessory extends too far from the
distal end of the attachment, or is not
properly centered in the attachment.
Remove and properly install the accessory.
If wobble persists, return the handpiece and
attachment to Stryker for repair.
A reamer attachment and/or accessory
is installed in the handpiece and the shift
collar is in the drill mode position.
Rotate the shift collar to the ream mode
position.
The handpiece is
noisy and/or vibrates.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece
experiences sporadic
electrical interference.
Electrical noise is present.
Turn off all electrical equipment not in use in
the operating room.
Relocate electrical equipment and/or
increase spatial distance between electrical
equipment.
Plug operating room equipment into different
operating room outlets.